Quinze-Vingts National Ophthalmology Hospital, Department of Ophthalmology, Paris, France.
Acta Ophthalmol. 2011 May;89(3):e225-30. doi: 10.1111/j.1755-3768.2010.01957.x. Epub 2010 Aug 25.
To analyse the reproducibility of corneal hysteresis (CH), corneal resistance factor (CRF), Goldmann-correlated intraocular pressure (IOPg) and corneal-compensated intraocular pressure (IOPcc) obtained with the ocular response analyzer (ORA).
This is a prospective study, nonmasked, of eight successive examinations with the ORA device in 60 normal eyes. Using 30 eyes (one eye per subject), the reproducibility was assessed by comparing the first series of four examinations to the second four and by calculating the within-subject coefficient of variation. The correlation and difference with the fellow eye were analysed, respectively.
The mean values were 10.7 ± 1.8 mmHg, CRF; 10.6 ± 1.6 mmHg, CH; 15.9 ± 3.9 mmHg, IOPg and 16.2 ± 3.7 mmHg, IOPcc. The reproducibility was significantly different for CRF (5.2 ± 5.9%), CH (7.3 ± 8.6%), IOPg (7.7 ± 6.7%) and IOPcc (10.1 ± 8.0%); p < 0.001. Considering the reproducibility, CRF correlated with CH (rs = 0.55; p < 0.001) and showed to be independent of IOPg and IOPcc. The score spread was best for CRF (2.6 ± 1.5 mmHg; 24.0%) compared to IOPg (4.3 ± 1.5 mmHg; 28.1%) and CH (3.1 ± 1.7 mmHg; 29.9%) and worst for IOPcc (5.5 ± 2.5 mmHg; 34.4%). The lowest difference with the fellow eye was observed for CRF (5.0%; p = 0.09). The correlation with the fellow eye was high, especially for IOPcc and CRF (rs > 0.9; p < 0.001) followed by IOPg and CH (rs > 0.8; p < 0.001).
The ORA device provides reproducible information on viscoelastic properties of the cornea in normal eyes notably CRF and CH. IOPcc was less reproducible. Four measurements per eye were necessary to reach a 10% precision and six for 5%.
分析眼反应分析仪(ORA)测量的角膜滞后(CH)、角膜阻力因子(CRF)、Goldmann 相关眼压(IOPg)和角膜补偿眼压(IOPcc)的可重复性。
这是一项前瞻性、非盲法研究,连续 8 次使用 ORA 设备对 60 只正常眼进行检查。使用 30 只眼(每只眼 1 只),通过比较前 4 次系列检查与后 4 次系列检查,计算受试者内变异系数,评估重复性。分析与对侧眼的相关性和差异。
平均数值为 10.7 ± 1.8mmHg、CRF;10.6 ± 1.6mmHg、CH;15.9 ± 3.9mmHg、IOPg 和 16.2 ± 3.7mmHg、IOPcc。CRF(5.2 ± 5.9%)、CH(7.3 ± 8.6%)、IOPg(7.7 ± 6.7%)和 IOPcc(10.1 ± 8.0%)的可重复性差异有统计学意义(p < 0.001)。考虑到可重复性,CRF 与 CH 相关(rs = 0.55;p < 0.001),且与 IOPg 和 IOPcc 无关。CRF 的评分分布最好(2.6 ± 1.5mmHg;24.0%),其次是 IOPg(4.3 ± 1.5mmHg;28.1%)和 CH(3.1 ± 1.7mmHg;29.9%),最差的是 IOPcc(5.5 ± 2.5mmHg;34.4%)。与对侧眼的差异最小的是 CRF(5.0%;p = 0.09)。与对侧眼的相关性很高,特别是 IOPcc 和 CRF(rs > 0.9;p < 0.001),其次是 IOPg 和 CH(rs > 0.8;p < 0.001)。
ORA 设备可提供正常眼角膜粘弹特性的可重复性信息,尤其是 CRF 和 CH。IOPcc 的可重复性较差。每只眼需要测量 4 次才能达到 10%的精度,6 次才能达到 5%的精度。