两种方向下磁共振频率分析测量的种植体稳定性指数比较:初始愈合期的前瞻性临床研究。

A comparison of implant stability quotients measured using magnetic resonance frequency analysis from two directions: a prospective clinical study during the initial healing period.

机构信息

Department of Oral & Maxillofacial Surgery, School of Dentistry, Seoul National University, Chongro-Gu, Seoul, Korea.

出版信息

Clin Oral Implants Res. 2010 Jun;21(6):591-7. doi: 10.1111/j.1600-0501.2009.01868.x. Epub 2010 Feb 1.

Abstract

OBJECTIVES

Given that the orientation of the transducer (mesiodistal or buccolingual) affects the data obtained from a piezoelectric resonance frequency analysis (RFA), this study evaluated whether it is necessary to use measurements taken in two different directions (mesiodistal and buccolingual) when using magnetic RFA to assess changes in the stiffness of dental implants.

MATERIALS AND METHODS

A prospective clinical trial was completed, in a total of 53 patients, on 71 non-submerged dental implants that were inserted to replace the unilateral loss of mandibular molars. All of the implants were of the same diameter (4.1 mm), length (10 mm), and collar height (2.8 mm). The implant stability quotient (ISQ) was measured during the surgical procedure, and at 4 and 10 weeks after surgery. Measurements were taken twice in each direction: in the buccolingual direction from the buccal side and in the mesiodistal direction from the mesial side. The average of two measurements in each direction was regarded as the representative ISQ of that direction. The higher and lower values of the two ISQs (buccolingual and mesiodistal) were also classified separately. In addition, the variation in ISQ was quantified by subtracting the lower value from the higher value, and the implants were classified into two groups according to this variation: one with ISQ variation of 3 or more and the other with a variation of <3.

RESULTS

There were no differences between the two ISQs when measured from different directions, but there were significant differences between the higher and lower values of the ISQs at each measurement point. A significant difference was also observed between the two ISQ variation groups in the pattern of change of the lower value for the period from immediately after surgery to 10 weeks after surgery.

CONCLUSION

Acquisition of two directional measurements and classification of the higher and lower values of the two directional ISQs may allow clinicians to detect patterns of change in ISQ that would not be identified if only one directional measurement were made.

摘要

目的

由于换能器的取向(颊舌向或近远中向)会影响从压电共振频率分析(RFA)获得的数据,因此本研究评估了当使用磁 RFA 评估牙种植体刚度变化时,是否有必要在两个不同方向(颊舌向和近远中向)进行测量。

材料和方法

本研究为前瞻性临床试验,共纳入 53 名患者的 71 个非潜入式牙种植体。这些种植体用于替代下颌磨牙单侧缺失。所有种植体的直径(4.1mm)、长度(10mm)和颈高(2.8mm)均相同。在手术过程中以及术后 4 周和 10 周测量种植体稳定性指数(ISQ)。在每个方向上均测量两次:从颊侧向颊舌向测量,从近中侧向近远中向测量。每个方向的两次测量的平均值被视为该方向的代表性 ISQ。也分别对两个 ISQ(颊舌向和近远中向)的较高值和较低值进行分类。此外,通过从较高值中减去较低值来量化 ISQ 的变化,并根据这种变化将种植体分为两组:ISQ 变化为 3 或更大的一组和 ISQ 变化小于 3 的一组。

结果

从不同方向测量时,两个 ISQ 之间没有差异,但在每个测量点的 ISQ 的较高值和较低值之间存在显著差异。在从手术即刻到术后 10 周的较低值变化模式方面,两个 ISQ 变化组之间也存在显著差异。

结论

获取两个方向的测量值并对两个方向 ISQ 的较高值和较低值进行分类,可能使临床医生能够检测到仅进行一个方向的测量时无法识别的 ISQ 变化模式。