影响种植体组织整合期间 Osstell mentor 评估的共振频率分析的因素:II. 种植体表面改性和种植体直径。

Factors influencing resonance frequency analysis assessed by Osstell mentor during implant tissue integration: II. Implant surface modifications and implant diameter.

机构信息

Peking University School, Hospital of Stomatology, The University of Hong Kong, Prince Philip Dental Hospital, Sai Ying Pun, Hong Kong, SAR, China.

出版信息

Clin Oral Implants Res. 2010 Jun;21(6):605-11. doi: 10.1111/j.1600-0501.2009.01909.x.

Abstract

OBJECTIVES

To monitor the development of the stability of Straumann tissue-level implants during the early phases of healing by resonance frequency analysis (RFA) and to determine the influence of implant surface modification and diameter.

MATERIAL AND METHODS

A total of twenty-five 10 mm length implants including 12 SLA RN v4.1 mm implants, eight SLActive RN v4.1 mm implants and five SLA WN v4.8 mm implants were placed. Implant stability quotient (ISQ) values were determined with Osstell mentor at baseline, 4 days, 1, 2, 3, 4, 6, 8 and 12 weeks post-surgery. ISQ values were compared between implant types using unpaired t-tests and longitudinally within implant types using paired t-tests.

RESULTS

During healing, ISQ decreased by 3-4 values after installation and reached the lowest values at 3 weeks. Following this, the ISQ values increased steadily for all implants and up to 12 weeks. No significant differences were noted over time. The longitudinal changes in the ISQ values showed the same patterns for SLA implants, SLActive implants and WB implants. At placement, the mean ISQ values were 72.6, 75.7 and 74.4, respectively. The mean lowest ISQ values, recorded at 3 weeks, were 69.9, 71.4 and 69.8, respectively. At 12 weeks, the mean ISQ values were 76.5, 78.8 and 77.8, respectively. The mean ISQ values at all observation periods did not differ significantly among the various types. Single ISQ values ranged from 55 to 84 during the entire healing period. Pocket probing depths of the implants ranged from 1 to 3 mm and bleeding on probing from 0 to 2 sites/implant post-surgically.

CONCLUSIONS

All ISQ values indicated the stability of Straumann implants over a 12-week healing period. All implants showed a slight decrease after installation, with the lowest ISQ values being reached at 3 weeks. ISQ values were restored 8 weeks post-surgically. It is recommended to monitor implant stability by RFA at 3 and 8 weeks post-surgically. However, neither implant surface modifications (SLActive) nor implant diameter were revealed by RFA.

摘要

目的

通过共振频率分析(RFA)监测 Straumann 组织水平种植体在早期愈合阶段的稳定性发展,并确定种植体表面改性和直径的影响。

材料与方法

共放置 25 个 10mm 长度的种植体,包括 12 个 SLA RN v4.1mm 种植体、8 个 SLActive RN v4.1mm 种植体和 5 个 SLA WN v4.8mm 种植体。在基线、术后 4 天、1、2、3、4、6、8 和 12 周时,使用 Osstell mentor 测定种植体稳定性指数(ISQ)值。使用未配对 t 检验比较不同种植体类型之间的 ISQ 值,并在种植体类型内进行纵向配对 t 检验。

结果

在愈合过程中,安装后 ISQ 值降低 3-4 个值,并在 3 周时达到最低值。此后,所有种植体的 ISQ 值均稳步升高,直到 12 周。随着时间的推移,没有发现显著差异。ISQ 值的纵向变化在 SLA 种植体、SLActive 种植体和 WB 种植体中呈现相同的模式。在放置时,ISQ 值的平均值分别为 72.6、75.7 和 74.4。在 3 周时记录到的最低 ISQ 值平均值分别为 69.9、71.4 和 69.8。在 12 周时,ISQ 值的平均值分别为 76.5、78.8 和 77.8。在所有观察期内,不同类型之间的 ISQ 值平均值没有显著差异。整个愈合期内,单个 ISQ 值范围为 55 至 84。种植体的牙周袋探诊深度范围为 1 至 3mm,术后探诊出血为 0 至 2 个部位/种植体。

结论

所有 ISQ 值均表明 Straumann 种植体在 12 周愈合期内的稳定性。所有种植体在安装后均略有下降,最低 ISQ 值在 3 周时达到。术后 8 周时 ISQ 值得到恢复。建议在术后 3 周和 8 周时通过 RFA 监测种植体稳定性。然而,RFA 既没有显示种植体表面改性(SLActive)也没有显示种植体直径的影响。