Lin Ming-Chih, Fu Yun-Ching, Jan Sheng-Ling, Ho Chi-Lin, Hwang Betau
Department of Pediatrics and Institute of Clinical Medicine, National Yang-Ming University.
Int Heart J. 2010 Jan;51(1):56-9. doi: 10.1536/ihj.51.56.
The stretched size of an atrial septal defect (ASD) is important for device selection during transcatheter closure. However, balloon sizing carries potential risks such as hypotension, bradycardia, or laceration of the atrial septum. The aim of the present study was to investigate the accuracy of the predicted stretched size of ASD by intracardiac echocardiography (ICE). From December 2004 to November 2007, 136 consecutive patients with single secundum type ASD undergoing transcatheter closure of their defect using the Amplatzer septal occluder under ICE guidance were enrolled for analysis. There were 43 males and 93 females. The age ranged from 2.2 to 79.1 years with a median age of 13.4 years. The body weight ranged from 12.1 to 93.2 kg with a median body weight of 45.8 kg. The stretched size of ASD measured by a sizing plate was considered as the standard. ASD sizes measured by ICE in bicaval and short-axis views predicted the stretched size by formulae derived from linear regressions. The predicted stretched sizes obtained using 2 formulae, 1.34 x radicalbicaval xshort axis (formula 1) and 1.22 x larger diameter (formula 2), exhibited good agreement with the standard stretched size with Kappa values of 0.91 and 0.90, respectively. The accuracy rate of predicted stretched sizes within 2 mm, 3 mm, and 4 mm range of the standard size were 32.8%, 45.4%, and 57.7% (formula 1) and 33.1%, 50%, and 63.2% (formula 2). The stretched size of ASD predicted by ICE exhibited good agreement with the standard stretched size. This prediction provides helpful information, especially if balloon sizing cannot be adequately performed.
房间隔缺损(ASD)的伸展大小对于经导管封堵术期间的器械选择很重要。然而,球囊测量大小存在潜在风险,如低血压、心动过缓或房间隔撕裂。本研究的目的是探讨心内超声心动图(ICE)预测ASD伸展大小的准确性。从2004年12月至2007年11月,连续纳入136例继发孔型ASD患者,这些患者在ICE引导下使用Amplatzer房间隔封堵器经导管封堵缺损并进行分析。其中男性43例,女性93例。年龄范围为2.2至79.1岁,中位年龄为13.4岁。体重范围为12.1至93.2 kg,中位体重为45.8 kg。用测量板测量的ASD伸展大小被视为标准。在双腔和短轴视图中通过ICE测量的ASD大小通过线性回归得出的公式预测伸展大小。使用两个公式获得的预测伸展大小,即1.34×双腔根号×短轴(公式1)和1.22×较大直径(公式2),与标准伸展大小具有良好的一致性,Kappa值分别为0.91和0.90。预测伸展大小在标准大小的2 mm、3 mm和4 mm范围内的准确率分别为32.8%、45.4%和57.7%(公式1)以及33.1%、50%和63.2%(公式2)。ICE预测的ASD伸展大小与标准伸展大小具有良好的一致性。这种预测提供了有用的信息,特别是在无法充分进行球囊测量大小时。