A novel experimental set-up for in operando studies of homogeneous catalyzed reactions under laboratory conditions has been developed and tested. It combines time-resolved X-ray absorption spectroscopy with UV/Vis spectroscopy. The reaction solution is stirred in a vessel and pumped in a circle by a peristaltic free gear-wheel through a measurement cell. The X-ray and UV/Vis beams probe the same sample volume of the cell orthogonally. Reactants can be added to the reaction mixture in the course of the measurements and a defined gas atmosphere can be adjusted up to a pressure of 10 bar. The in situ reduction of cerium(IV) ammonium nitrate to cerium(III) by isopropanol is studied as a test reaction with quick-XANES and UV/Vis measurements with a time resolution of 60 s and 1 s, respectively.
已经开发并测试了一种新的实验装置,用于在实验室条件下对均相催化反应进行原位研究。它将时间分辨 X 射线吸收光谱与紫外/可见光谱相结合。反应溶液在容器中搅拌,并通过蠕动自由齿轮泵以循环方式通过测量池。X 射线和紫外/可见光束正交探测同一测量池的样品体积。在测量过程中可以向反应混合物中添加反应物,并可以调节一定的气体气氛,最高压力可达 10 巴。异丙醇将硝酸铈(IV)铵原位还原为铈(III)的反应作为测试反应,快速 XANES 和紫外/可见测量的时间分辨率分别为 60 s 和 1 s。