Herget W F, Jahnke J A, Burch D E, Gryvnak D A
Appl Opt. 1976 May 1;15(5):1222-8. doi: 10.1364/AO.15.001222.
An ir analyzer employing gas-filter correlation techniques has been designed and constructed to measure the concentrations of CO, NO, SO(2), HCl, and HF in the stacks or ducts of stationary pollutant sources. Use of a retroreflector allows the stack to be double passed, and no sample is extracted. For each gas, small interchangeable fixed-position grating polychromators are used as narrow (~1.5-cm(-1)) multiband spectral filters with the bands corresponding to locations of selected absorption lines. The approximate useful ranges (in parts per million-meters) over which this analyzer operates are 10-4000 for NO, 10-1500 for CO, 50-40,000 for SO(2), 10-2000 for HC1, and 5-200 for HF. The discrimination against other gases and particulates is excellent. The analyzer has been tested in the laboratory and on a variety of pollutant sources.