Benedetti P A, Bianchini G, Chiti G
Appl Opt. 1976 Oct 1;15(10):2554-8. doi: 10.1364/AO.15.002554.
The poor speed performance or the limited optical accuracy of the scanning devices currently employed in microspectroscopy can be substantially improved, for high resolution work, using a method based on the displacement of the condenser by means of an electrodynamic technique. The unit described, for work in the visible range, features random addressing capability for both X and Y scanning axes and focusing by means of arbitrary driving signals. An area up to 500 x 500 microm(2) can be explored with an accuracy of 0.15 microm approximately, while the position settling-time is less than 6 msec. The fast and precise operation is particularly valuable in dual-beam measurements on photosensitive and living samples as well as in more complicated computer assisted experiments.
目前用于显微光谱学的扫描设备速度性能不佳或光学精度有限,对于高分辨率工作而言,通过一种基于电动技术移动聚光镜的方法,其性能可得到显著改善。所描述的该装置用于可见光范围工作,具有X和Y扫描轴的随机寻址能力,并通过任意驱动信号进行聚焦。面积达500×500微米²的区域可被探测,精度约为0.15微米,而位置稳定时间小于6毫秒。这种快速且精确的操作在对光敏和活体样本的双光束测量以及更复杂的计算机辅助实验中尤其有价值。