Ginsel Bastiaan L, Banks Scott, Verdonschot Nico, Hodge W Andrew
Biomotion Foundation, West Palm Beach, Florida, USA.
Acta Orthop Belg. 2009 Dec;75(6):801-7.
Achieving a large range of motion (ROM) is a much-desired clinical outcome after total knee arthroplasty (TKA), especially in Asian and Middle Eastern cultures. TKA design plays an important role in providing the post-operative ROM. This study investigated the kinematics of a new high-flexion posterior cruciate ligament retaining total knee replacement, featuring an enlarged posterior condylar offset and a more conforming tibiofemoral articulation. Two flexion activities were compared to determine which provides higher flexion kinematics. Sixteen North American patients with 20 total knee implants were studied using fluoroscopy and shape matching techniques. Maximum skeletal flexion during a lunge activity averaged 120 degrees +/- 11 degrees, with 11 degrees +/- 4 degrees tibial internal rotation. Kneeling activities showed 11 degrees greater average maximum skeletal flexion (131 degrees +/- 13 degrees, p < 0.05) and 1 degrees less tibial internal rotation (10 degrees +/- 4 degrees, p > 0.05) than lunge activities. We conclude that specific knee implant design features can facilitate high flexion in fixed-bearing cruciate retaining TKA, and that kneeling activities provide higher flexion than lunge activities.
全膝关节置换术(TKA)后实现大范围活动度(ROM)是一个备受期待的临床结果,在亚洲和中东文化背景下尤其如此。TKA的设计在提供术后活动度方面起着重要作用。本研究调查了一种新型高屈曲后交叉韧带保留型全膝关节置换术的运动学,其特点是后髁偏移增大以及胫股关节面更贴合。比较了两种屈曲活动以确定哪种能提供更高的屈曲运动学。使用荧光透视和形状匹配技术对16名北美患者的20个全膝关节植入物进行了研究。弓步活动期间的最大骨骼屈曲平均为120度±11度,伴有11度±4度的胫骨内旋。跪姿活动显示平均最大骨骼屈曲比弓步活动大11度(131度±13度,p<0.05),胫骨内旋比弓步活动少1度(10度±4度,p>0.05)。我们得出结论,特定的膝关节植入物设计特征可促进固定平台后交叉韧带保留型TKA的高屈曲,并且跪姿活动比弓步活动能提供更高的屈曲度。