Thilander-Klang Anne, Ledenius Kerstin, Hansson Jonny, Sund Patrik, Båth Magnus
Department of Radiation Physics, Institute of Clinical Sciences, The Sahlgrenska Academy, University of Gothenburg, SE-41345 Gothenburg, Sweden.
Radiat Prot Dosimetry. 2010 Apr-May;139(1-3):449-54. doi: 10.1093/rpd/ncq069. Epub 2010 Feb 22.
The purpose of the present work was to investigate the reliability of subjective assessments of the low-contrast visibility in constancy control of computed tomography (CT). Axial CT images of a low-contrast phantom were acquired on an 8-slice multi-detector CT scanner at nine tube current settings ranging from 75 to 440 mA. Five medical physicists assessed the visibility of the low-contrast details in two sessions. In the first session, containing 54 images, the visibility was rated on an absolute scale by determining the number of visible details in each contrast group in each image. In the second session, 180 image pairs were presented to the observers with the task of determining if the two images had been acquired under identical conditions or not. In the absolute session, both the intra- and inter-observer variabilities were high. In the relative session, the variability was smaller, but an exposure difference of 50 % was needed for all observers to correctly identify a change in all cases. In conclusion, the present study indicates that subjective assessments of the low-contrast visibility in constancy control of CT are not reliable.
本研究的目的是调查在计算机断层扫描(CT)的稳定性控制中,对低对比度可见性进行主观评估的可靠性。在一台8层多探测器CT扫描仪上,于75至440 mA的九个管电流设置下采集低对比度体模的轴向CT图像。五名医学物理学家在两个阶段评估了低对比度细节的可见性。在第一阶段,包含54幅图像,通过确定每幅图像中每个对比度组中可见细节的数量,以绝对尺度对可见性进行评级。在第二阶段,向观察者呈现180对图像,任务是确定这两幅图像是否在相同条件下采集。在绝对尺度阶段,观察者内部和观察者之间的变异性都很高。在相对阶段,变异性较小,但所有观察者都需要50%的曝光差异才能在所有情况下正确识别变化。总之,本研究表明,在CT稳定性控制中对低对比度可见性的主观评估不可靠。