Suppr超能文献

使用掺铒钇铝石榴石激光和掺铒铬钇钪镓石榴石激光对根管根尖三分之一处使用的改良牙髓激光头的热效应。

Thermal effects from modified endodontic laser tips used in the apical third of root canals with erbium-doped yttrium aluminium garnet and erbium, chromium-doped yttrium scandium gallium garnet lasers.

作者信息

George Roy, Walsh Laurence J

机构信息

School of Dentistry, University of Queensland, Brisbane, Australia.

出版信息

Photomed Laser Surg. 2010 Apr;28(2):161-5. doi: 10.1089/pho.2008.2423.

Abstract

OBJECTIVE

To evaluate the temperature changes occurring on the apical third of root surfaces when erbium-doped yttrium aluminium garnet (Er:YAG) and erbium, chromium-doped yttrium scandium gallium garnet (Er,Cr:YSGG) laser energy was delivered with a tube etched, laterally emitting conical tip and a conventional bare design optical fiber tip.

BACKGROUND DATA

Thermal effects of root canal laser treatments on periodontal ligament cells and alveolar bone are of concern in terms of safety.

MATERIALS AND METHODS

A total of 64 single-rooted extracted teeth were prepared 1 mm short of the working length using rotary nickel-titanium Pro-Taper files to an apical size corresponding to a F5 Pro-Taper instrument. A thermocouple located 2 mm from the apex was used to record temperature changes arising from delivery of laser energy through laterally emitting conical tips or plain tips, using an Er:YAG or Er,Cr:YSGG laser.

RESULTS

For the Er:YAG and Er,Cr:YSGG systems, conical fibers showed greater lateral emissions (452 + 69% and 443 + 64%) and corresponding lower forward emissions (48 + 5% and 49 + 5%) than conventional plain-fiber tips. All four combinations of laser system and fiber design elicited temperature increases less than 2.5 degrees C during lasing. The use of water irrigation attenuated completely the thermal effects of individual lasing cycles.

CONCLUSIONS

Laterally emitting conical fiber tips can be used safely under defined conditions for intracanal irradiation without harmful thermal effects on the periodontal apparatus.

摘要

目的

评估掺铒钇铝石榴石(Er:YAG)激光和掺铒铬钇钪镓石榴石(Er,Cr:YSGG)激光通过蚀刻管的侧向发射锥形尖端和传统裸光纤尖端传递能量时,牙根表面根尖三分之一处的温度变化。

背景数据

根管激光治疗对牙周膜细胞和牙槽骨的热效应在安全性方面备受关注。

材料与方法

使用旋转镍钛Pro-Taper锉将64颗单根离体牙预备至比工作长度短1mm,根尖尺寸对应F5 Pro-Taper器械。使用距根尖2mm处的热电偶记录通过侧向发射锥形尖端或普通尖端传递激光能量时产生的温度变化,激光采用Er:YAG或Er,Cr:YSGG。

结果

对于Er:YAG和Er,Cr:YSGG系统,与传统普通光纤尖端相比,锥形光纤显示出更大的侧向发射(分别为452 + 69%和443 + 64%)以及相应更低的向前发射(分别为48 + 5%和49 + 5%)。激光系统和光纤设计的所有四种组合在激光照射期间引起的温度升高均小于2.5℃。使用水冲洗可完全减弱单个激光照射周期的热效应。

结论

在特定条件下,侧向发射锥形光纤尖端可安全用于根管内照射,而不会对牙周组织产生有害热效应。