J Cataract Refract Surg. 2010 Mar;36(3):477-82. doi: 10.1016/j.jcrs.2009.09.040.
To compare real-time intraocular pressure (IOP) between laser in situ keratomileusis (LASIK) and epithelial LASIK (epi-LASIK) in porcine eyes during flap creation using a microkeratome or an epikeratome, respectively.
Vissum Madrid, Madrid, Spain.
In this prospective study, a Moria microkeratome was used in 1 eye (LASIK group) and an Epi-K epikeratome in the other eye (epi-LASIK group) to create a lamellar corneal flap and an epithelial flap, respectively, in freshly enucleated porcine eyes. The IOP changes during the procedures were recorded by direct cannulation using a reusable blood pressure transducer connected to the anterior chamber.
Each group comprised 17 eyes. In the LASIK group, the mean IOP was 113.65 mm Hg +/- 10.78 (SD) during suctioning and 112.35 +/- 11.51 mm Hg during cutting phases. The mean duration of the phases was 9.00 +/- 1.46 seconds and 6.06 +/- 1.14 seconds, respectively. In the epi-LASIK group, the mean IOP was 92.57 +/- 20.86 mm Hg during suctioning, 82.09 +/- 20 mm Hg during cutting, and 67.28 +/- 13.49 during low vacuum. The mean duration of the phases was 25.88 +/- 1.96 seconds, 33.82 +/- 2.81 seconds, and 29.71 +/- 3.29 seconds, respectively. The IOP values were significantly different between the 2 groups (all comparisons P<.05).
Real-time IOP measured during suctioning and flap creation by direct cannulation of the anterior chamber in freshly enucleated porcine eyes showed a significant increase in IOP during LASIK and epi-LASIK; the increase was lower in the epi-LASIK group.
比较微角膜刀和角膜上皮刀分别行 LASIK 和上皮下角膜切除术(epi-LASIK)时猪眼瓣制作过程中实时眼内压(IOP)的变化。
西班牙马德里 Vissum 公司。
本前瞻性研究中,17 只新鲜猪眼分别使用 Moria 微角膜刀(LASIK 组)和 Epi-K 角膜上皮刀(epi-LASIK 组)制作角膜瓣和上皮瓣。通过连接到前房的可重复使用的血压换能器直接插管记录手术过程中的 IOP 变化。
LASIK 组和 epi-LASIK 组各 17 只眼。LASIK 组中,吸引阶段的平均 IOP 为 113.65 ± 10.78mmHg,切割阶段为 112.35 ± 11.51mmHg。各阶段的平均持续时间分别为 9.00 ± 1.46 秒和 6.06 ± 1.14 秒。epi-LASIK 组中,吸引阶段的平均 IOP 为 92.57 ± 20.86mmHg,切割阶段为 82.09 ± 20mmHg,低真空阶段为 67.28 ± 13.49mmHg。各阶段的平均持续时间分别为 25.88 ± 1.96 秒、33.82 ± 2.81 秒和 29.71 ± 3.29 秒。2 组间 IOP 值差异均有统计学意义(P<0.05)。
直接插管测量新鲜猪眼前房内 LASIK 和 epi-LASIK 时的实时 IOP 显示,两种手术均会导致 IOP 显著升高;epi-LASIK 组升高幅度较低。