John A. Moran Eye Center, University of Utah, Salt Lake City, Utah 84132, USA.
J Cataract Refract Surg. 2010 Mar;36(3):512-6. doi: 10.1016/j.jcrs.2009.08.040.
We studied an enucleated postmortem eye from an 82-year-old white donor who had been implanted with a Pannu "universal" intraocular lens (IOL) in the anterior chamber approximately 20 years earlier. This IOL has design features characteristic of a 1-piece, C-loop posterior chamber IOL. Magnetic resonance imaging showed a relatively well-centered IOL in the anterior chamber with haptics impinging on the iris. Gross and light microscopic analyses of the eye and the IOL showed peripheral anterior synechiae enclaving one haptic, areas of angle widening, significant attenuation of the corneal endothelium, multiple areas of iris trauma secondary to optic and haptic iris abrasion, large areas of pigment dispersion in the angle, diffuse pigment accumulation within the anterior chamber, and attenuation of the ganglion cell layer. The histopathological findings were consistent with glaucoma and chronic inflammation.
我们研究了一只来自 82 岁白人供体的眼标本,该供体在大约 20 年前被植入了 Pannu“通用”前房型人工晶状体(IOL)。该 IOL 具有 1 件式、C 袢后房型 IOL 的设计特点。磁共振成像显示在前房内相对居中的 IOL,襻撞击虹膜。对眼睛和 IOL 的大体和光镜分析显示,周边前粘连包绕着一个襻,房角变宽,角膜内皮显著衰减,多处虹膜创伤继发于光学襻和虹膜擦伤,角部有大量色素播散,前房内弥漫性色素沉着,以及节细胞层衰减。组织病理学发现与青光眼和慢性炎症一致。