Abstract

BACKGROUND

Because strict glucose control is important for reducing the complications of diabetes, the self-monitoring of blood glucose is one of the fundamental treatment modalities. Many glucometers have been developed. In the present study, we evaluated a new glucometer: GlucoDr Plus (Allmedicus, Anyang, Gyeonggi-do, Republic of Korea).

METHODS

The evaluation was performed based on Clinical and Laboratory Standards Institute guidelines. Interferences by ascorbic acid, uric acid, maltose, and acetaminophen were examined, and the performance of the unit was compared to those of six other glucometers. The effects of hematocrit, of oxygen partial pressure (PaO(2)), and of multiple users were also evaluated.

RESULTS

Within-run, between-run, between-day, and total imprecision (coefficients of variation) were 0.99-4.98%. Satisfactory linearity was found for glucose concentrations of 32.5-786.5 mg/dL (R(2) = 0.9985). A comparison with the reference laboratory method showed close concordance over the entire range of concentrations evaluated (R(2) = 0.9869). No significant effects were noted due to added interferents, hematocrit, and PaO(2).

CONCLUSIONS

The GlucoDr Plus showed acceptable performance in terms of precision and linearity. It was minimally affected by various interferents. GlucoDr Plus is suitable for the self-monitoring of blood glucose by patients with diabetes.

摘要

背景

由于严格的血糖控制对于减少糖尿病并发症很重要,因此自我监测血糖是基本的治疗方法之一。已经开发出许多血糖仪。在本研究中,我们评估了一种新的血糖仪:GlucoDr Plus(Allmedicus,安养,京畿道,大韩民国)。

方法

根据临床和实验室标准协会的指南进行评估。检查了抗坏血酸、尿酸、麦芽糖和对乙酰氨基酚的干扰,并将该仪器的性能与其他六台血糖仪进行了比较。还评估了红细胞压积、氧分压(PaO2)和多位使用者的影响。

结果

批内、批间、日间和总不精密度(变异系数)分别为 0.99-4.98%。在 32.5-786.5mg/dL 的葡萄糖浓度范围内发现了令人满意的线性关系(R2=0.9985)。与参考实验室方法的比较表明,在整个评估浓度范围内具有很好的一致性(R2=0.9869)。添加干扰物、红细胞压积和 PaO2 没有显著影响。

结论

GlucoDr Plus 在精密度和线性方面表现出可接受的性能。它受各种干扰物的影响很小。GlucoDr Plus 适合糖尿病患者自我监测血糖。