Philips Corporate Technologies, Eindhoven, The Netherlands.
Clin Chim Acta. 2010 Jun 3;411(11-12):868-73. doi: 10.1016/j.cca.2010.03.001. Epub 2010 Mar 6.
We present a prototype handheld device based on a newly developed optomagnetic technology for the sensitive detection of cardiac troponin I (cTnI) in a finger-prick blood sample with a turnaround time of 5 min.
The test was completed in a compact plastic disposable with on-board dry reagents and superparamagnetic nanoparticles. In our one-step assay, all reaction processes were precisely controlled using electromagnets positioned above and below the disposable. Nanoparticle labels (500 nm) bound to the sensor surface via a sandwich immunoassay were detected using the optical technique of frustrated total internal reflection.
A calibration function measured in plasma demonstrates a limit of detection (mean of blank plus 3-fold the standard deviation) of 0.03 ng/mL cTnI. A linear regression analysis of the region 0.03-6.5 ng/mL yields a slope of 37+/-4, and a linear correlation coefficient of R2=0.98. The measuring range could be extended substantially to 100 ng/mL by simultaneously imaging a second spot with a lower antibody concentration.
The combination of magnetic particles and their fine actuation with electromagnets permits the rapid and sensitive detection of cTnI. Because of the potential high analytical performance and ease-of-use of the test, it is well suited for demanding point-of-care diagnostic applications.
我们展示了一种基于新开发的磁光技术的原型手持式设备,用于在 5 分钟内从指尖采血样中灵敏检测心肌肌钙蛋白 I(cTnI)。
该测试在带有内置干试剂和超顺磁纳米粒子的紧凑型塑料一次性装置中完成。在我们的一步法测定中,所有反应过程都通过位于一次性装置上方和下方的电磁铁精确控制。通过夹心免疫测定与传感器表面结合的纳米粒子标记物(500nm)使用全内反射的光学技术进行检测。
在等离子体中测量的校准函数表明 cTnI 的检测限(空白的平均值加上 3 倍标准差)为 0.03ng/mL。0.03-6.5ng/mL 区域的线性回归分析得到斜率为 37+/-4,线性相关系数 R2=0.98。通过同时对具有较低抗体浓度的第二个斑点进行成像,可以将测量范围大大扩展到 100ng/mL。
磁性粒子及其与电磁铁的精细驱动相结合,可实现 cTnI 的快速灵敏检测。由于该测试具有潜在的高分析性能和易用性,非常适合苛刻的即时诊断应用。