McCoach Robert, Kunselman Allen, El-Banayosy Aly, Undar Akif
Penn State Hershey Pediatric Cardiovascular Research Center, Department of Pediatrics, Penn State College of Medicine, Penn State Children's Hospital, Hershey, Pennsylvania 17033-0850, USA.
Perfusion. 2010 Mar;25(2):71-6. doi: 10.1177/0267659110365366. Epub 2010 Mar 8.
Centrifugal blood pumps have been widely adopted in conventional adult cardiopulmonary bypass and circulatory assist procedures. Different brands of centrifugal blood pumps incorporate distinct designs which affect pump performance. In this adult extracorporeal life support (ECLS) model, the performances of two brands of centrifugal blood pump (RotaFlow blood pump and CentriMag blood pump) were compared.
The simulated adult ECLS circuit used in this study included a centrifugal blood pump, Quadrox D membrane oxygenator and Sorin adult ECLS tubing package. A Sorin Cardiovascular(R) VVR(R) 4000i venous reservoir (Sorin S.p.A., Milan, Italy) with a Hoffman clamp served as a pseudo-patient. The circuit was primed with 900ml heparinized human packed red blood cells and 300ml lactated Ringer's solution (total volume 1200 ml, corrected hematocrit 40%). Trials were conducted at normothermia (36 degrees C). Performance, including circuit pressure and flow rate, was measured for every setting analyzed.
The shut-off pressure of the RotaFlow was higher than the CentriMag at all measurement points given the same rotation speed (p < 0.0001). The shut-off pressure differential between the two centrifugal blood pumps was significant and increased given higher rotation speeds (p < 0.0001). The RotaFlow blood pump has higher maximal flow rate (9.08 +/- 0.01L/min) compared with the CentriMag blood pump (8.37 +/- 0.02L/min) (p < 0.0001). The blood flow rate differential between the two pumps when measured at the same revolutions per minute (RPM) ranged from 1.64L/min to 1.73L/min.
The results obtained in this experiment demonstrate that the RotaFlow has a higher shut-off pressure (less retrograde flow) and maximal blood flow rate than the CentriMag blood pump. Findings support the conclusion that the RotaFlow disposable pump head has a better mechanical performance than the CentriMag. In addition, the RotaFlow disposable pump is 20-30 times less expensive than the CentriMag.
离心式血泵已广泛应用于传统的成人体外循环和循环辅助手术中。不同品牌的离心式血泵采用了不同的设计,这会影响泵的性能。在这个成人体外生命支持(ECLS)模型中,比较了两个品牌的离心式血泵(RotaFlow血泵和CentriMag血泵)的性能。
本研究中使用的模拟成人体外生命支持回路包括一个离心式血泵、Quadrox D膜式氧合器和索林成人ECLS管路套件。一个带有霍夫曼夹的索林心血管(R)VVR(R)4000i静脉贮血器(意大利米兰索林集团股份公司)作为模拟患者。用900毫升肝素化人浓缩红细胞和300毫升乳酸林格氏液(总体积1200毫升,校正血细胞比容40%)对回路进行预充。在常温(36摄氏度)下进行试验。对每个分析设置测量包括回路压力和流速在内的性能。
在相同转速下,RotaFlow的关闭压力在所有测量点均高于CentriMag(p < 0.0001)。两个离心式血泵之间的关闭压力差显著,且随着转速升高而增大(p < 0.0001)。与CentriMag血泵(8.37 +/- 0.02升/分钟)相比,RotaFlow血泵具有更高的最大流速(9.08 +/- 0.01升/分钟)(p < 0.0001)。在相同每分钟转数(RPM)下测量时,两个泵之间的血流速度差在1.64升/分钟至1.73升/分钟之间。
本实验获得的结果表明,RotaFlow比CentriMag血泵具有更高的关闭压力(逆向血流更少)和最大血流速度。研究结果支持以下结论:RotaFlow一次性泵头的机械性能优于CentriMag。此外,RotaFlow一次性泵比CentriMag便宜20至30倍。