在小儿体外生命支持模型中对三种离心式血泵避免逆行血流的最低转速进行的一项研究。

An investigational study of minimum rotational pump speed to avoid retrograde flow in three centrifugal blood pumps in a pediatric extracorporeal life support model.

作者信息

Clark Joseph B, Guan Yulong, McCoach Robert, Kunselman Allen R, Myers John L, Undar Akif

机构信息

Department of Pediatrics, Penn State Hershey Pediatric Cardiovascular Research Center, Penn State Hershey, 500 University Drive, Hershey, PA 17033, USA.

出版信息

Perfusion. 2011 May;26(3):185-90. doi: 10.1177/0267659110394712. Epub 2011 Jan 12.

Abstract

During extracorporeal life support with centrifugal blood pumps, retrograde pump flow may occur when the pump revolutions decrease below a critical value determined by the circuit resistance and the characteristics of the pump. We created a laboratory model to evaluate the occurrence of retrograde flow in each of three centrifugal blood pumps: the Rotaflow, the CentriMag, and the Bio-Medicus BP-50. At simulated patient pressures of 60, 80, and 100 mmHg, each pump was evaluated at speeds from 1000 to 2200 rpm and flow rates were measured. Retrograde flow occurred at low revolution speeds in all three centrifugal pumps. The Bio-Medicus pump was the least likely to demonstrate retrograde flow at low speeds, followed by the Rotaflow pump. The CentriMag pump showed the earliest transition to retrograde flow, as well as the highest degree of retrograde flow. At every pump speed evaluated, the Bio-Medicus pump delivered the highest antegrade flow and the CentriMag pump delivered the least.

摘要

在使用离心血泵进行体外生命支持期间,当泵的转速降至由回路阻力和泵的特性所确定的临界值以下时,可能会出现逆行泵流。我们创建了一个实验室模型,以评估三种离心血泵(Rotaflow、CentriMag和Bio-Medicus BP-50)中每种泵逆行流的发生情况。在模拟患者压力为60、80和100 mmHg时,对每个泵在1000至2200 rpm的转速下进行评估,并测量流速。在所有三种离心泵中,低转速时均会出现逆行流。Bio-Medicus泵在低速时最不容易出现逆行流,其次是Rotaflow泵。CentriMag泵最早出现向逆行流的转变,并且逆行流程度最高。在评估的每个泵转速下,Bio-Medicus泵输送的顺行流最高,而CentriMag泵输送的顺行流最少。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验