基于体内心脏衰减的 PET 图像局部统计噪声的评估。

Estimation of local statistical noise in PET images induced by attenuation inside the body.

机构信息

National Cancer Center Hospital East, Kashiwa, Japan.

出版信息

Ann Nucl Med. 2010 Apr;24(3):197-205. doi: 10.1007/s12149-010-0355-0. Epub 2010 Mar 16.

Abstract

OBJECTIVES

In positron emission tomography/computed tomography (PET/CT) examinations, the standardized uptake value (SUV) is a commonly used index to evaluate the activity of cancer. The precision of SUV is directly affected by the local statistical noise in PET images because SUV is calculated based on the counts on PET image data. The purpose of this study was to estimate the local statistical noise in the PET image caused by attenuation of annihilation photons inside the body.

METHODS

To estimate the local statistical noise, we defined the number proportional to the standard deviation of a voxel caused by attenuation using CT image data. The conversion table between the number obtained from the CT image data and the actually measured standard deviation of a voxel was made by performing experiments with different forms or sizes of phantoms and scattering bodies. We compared the standard deviation of a voxel estimated by the method we proposed with actually measured ones.

RESULTS

The standard deviation of a voxel in the PET image varied according to the size of the phantom. The effect of attenuation on the standard deviation of a voxel was more significant in the 3D mode than in the 2D mode. The correlation coefficients between the estimated standard deviation and actually measured standard deviation were 0.98 and 0.99 in the 2D mode and 3D mode, respectively.

CONCLUSIONS

Our algorithm favorably estimated the local statistical noise in PET image data using CT image data, regardless of object size, although the consideration was limited for phantoms with homogeneous interiors.

摘要

目的

在正电子发射断层扫描/计算机断层扫描(PET/CT)检查中,标准化摄取值(SUV)是评估癌症活性的常用指标。SUV 的精度直接受到 PET 图像中的局部统计噪声的影响,因为 SUV 是基于 PET 图像数据的计数计算的。本研究的目的是估计体内湮灭光子衰减引起的 PET 图像中的局部统计噪声。

方法

为了估计局部统计噪声,我们使用 CT 图像数据定义了与体素衰减引起的标准差成正比的数字。通过使用不同形状或大小的体模和散射体进行实验,制作了从 CT 图像数据获得的数字与实际测量的体素标准差之间的转换表。我们比较了我们提出的方法估计的体素标准差与实际测量的标准差。

结果

PET 图像中体素的标准差随体模的大小而变化。衰减对体素标准差的影响在 3D 模式下比在 2D 模式下更为显著。在 2D 模式和 3D 模式下,估计的标准差与实际测量的标准差之间的相关系数分别为 0.98 和 0.99。

结论

无论物体大小如何,我们的算法都可以使用 CT 图像数据有利地估计 PET 图像数据中的局部统计噪声,尽管对于内部均匀的体模的考虑是有限的。