Boivin L P
Appl Opt. 1980 Aug 15;19(16):2771-80. doi: 10.1364/AO.19.002771.
A method for calibrating incandescent lamps for spectral irradiance by means of absolute radiometers is described in which a secondary radiometer is calibrated spectrally against absolute radiometers and then used in conjunction with a series of filters to calibrate the lamps. Considering both narrowband and wideband filters, an extensive mathematical error analysis is performed. The use of narrowband filters (20-25-nm halfwidth) is found to be advantageous because very little information is required on the spectral distribution of the lamp being measured. The most serious source of error is a wavelength shift in the measured spectral transmittances of the filters, especially at shorter avelengths; for example, at 400 nm, a wavelength shift error of 1 nm can cause an error approaching 3%. It is estimated that the accuracy of spectral irradiance measurements made using the method described here will vary between +/-1 and +/-0.5% from ~350 to 800 nm. Measurements on 500-W quartz-bromine spectral irradiance standards are described. With such lamps, only four or five narrowband filters are required to cover the spectral range from the near UV to the near IR. The measured and calibration values agreed to ~ +/-0.5% on average with a maximum difference of ~1%.
本文描述了一种利用绝对辐射计校准白炽灯光谱辐照度的方法,其中,二级辐射计先相对于绝对辐射计进行光谱校准,然后与一系列滤光片配合使用来校准灯泡。考虑到窄带和宽带滤光片,进行了广泛的数学误差分析。发现使用窄带滤光片(半高宽为20 - 25纳米)具有优势,因为对于被测灯泡的光谱分布所需信息极少。最严重的误差源是滤光片测量光谱透过率中的波长偏移,尤其是在较短波长处;例如,在400纳米处,1纳米的波长偏移误差可能导致接近3%的误差。据估计,使用本文所述方法进行光谱辐照度测量的精度在约350至800纳米范围内将在±1%至±0.5%之间变化。描述了对500瓦石英溴光谱辐照度标准的测量。对于此类灯泡,仅需四或五个窄带滤光片即可覆盖从近紫外到近红外的光谱范围。测量值与校准值平均相差约±0.5%,最大差值约为1%。