Division of Advanced Prosthetic Dentistry, Tohoku University Graduate School of Dentistry, Sendai, Japan.
Int J Prosthodont. 2010 Jan-Feb;23(1):38-41.
The retentive force of a wire clasp composed of a b-type Ti-6Mo-4Sn alloy (b-Ti alloy) with a low Young modulus was evaluated using a piezoelectric transducer to determine the appropriate undercut for removable partial dentures. There were no significant differences in retentive force between a b-Ti alloy wire with a 0.50-mm undercut and a cobalt-chromium alloy (Co-Cr alloy) wire with a 0.25-mm undercut, or between a b-Ti alloy wire with a 0.75-mm undercut and a Co-Cr alloy wire with a 0.50-mm undercut. The b-Ti alloy wire may be applicable for abutment teeth with a large number of undercuts. Int J Prosthodont 2010;23:38-41.
采用压电传感器评估了由低杨氏模量的 b 型 Ti-6Mo-4Sn 合金(b-Ti 合金)制成的卡环的固位力,以确定可摘局部义齿的合适的就位道。0.50mm 深的 b-Ti 合金丝与 0.25mm 深的钴铬合金(Co-Cr 合金)丝之间、0.75mm 深的 b-Ti 合金丝与 0.50mm 深的 Co-Cr 合金丝之间的固位力没有显著差异。b-Ti 合金丝可能适用于具有大量就位道的基牙。国际口腔修复学杂志 2010;23:38-41.