Kato A, Yoshimine T, Hayakawa T, Tomita Y, Ikeda T, Mitomo M, Harada K, Mogami H
Department of Neurosurgery, Osaka University Medical School.
No Shinkei Geka. 1991 Feb;19(2):137-42.
We have developed a computer assisted neurosurgical navigation system, which displays the point of ongoing surgical manipulation on the CT or MR images. The system utilizes magnetic field modulation technology for the detection of the 3 dimensional position and orientation angles of the probe, a resin pointer or suction tube with a small magnetic field sensor. Four fiducial markers on the scalp were used to translate the spatial data of the probe onto the CT/MRI images. This frameless and armless navigator with a spatial accuracy of 4mm, introduces CT/MRI-stereotaxy to conventional open neurosurgery without limiting the operative field or interfering with surgical procedures.
我们开发了一种计算机辅助神经外科导航系统,该系统可在CT或MR图像上显示正在进行手术操作的位置。该系统利用磁场调制技术来检测带有小型磁场传感器的探针、树脂指针或吸管的三维位置和方向角度。头皮上的四个基准标记用于将探针的空间数据转换到CT/MRI图像上。这种空间精度为4毫米的无框架、无臂导航仪将CT/MRI立体定向技术引入传统的开放性神经外科手术中,而不会限制手术视野或干扰手术操作。