Department of Cardiology, Maria Middelares Hospital, Kortrijksesteenweg 1026, B-9000 Ghent, Belgium.
Europace. 2010 Aug;12(8):1195-6. doi: 10.1093/europace/euq091. Epub 2010 Mar 27.
Arrhythmogenic right ventricular cardiomyopathy (ARVC) is a disease where therapy with an internal cardioverter defibrillator (ICD) is frequently indicated, but the use of these devices is often challenged because of sensing dysfunction with low right ventricular sensing amplitude. We present the case of an ARVC patient in whom ICD sensing difficulties were overcome by the implantation of a coronary sinus lead for pacing and sensing.
致心律失常性右室心肌病 (ARVC) 是一种常需要植入体内除颤器 (ICD) 进行治疗的疾病,但由于右心室感知幅度低导致感知功能障碍,这些设备的使用常常受到挑战。我们介绍了一例 ARVC 患者,通过植入冠状窦导线进行起搏和感知,解决了 ICD 感知困难的问题。