Suppr超能文献

实时接触力曲线下面积(力-时间积分)可预测体外收缩模型中的射频消融灶大小。

Area under the real-time contact force curve (force-time integral) predicts radiofrequency lesion size in an in vitro contractile model.

机构信息

Electrophysiology Unit, Cardiology Service, Hôpitaux Universitaires de Genéve, Geneva, Switzerland.

出版信息

J Cardiovasc Electrophysiol. 2010 Sep;21(9):1038-43. doi: 10.1111/j.1540-8167.2010.01750.x.

Abstract

INTRODUCTION

Electrode tissue contact, radiofrequency (RF) power and duration are major determinants of RF lesion size. Since contact forces (CF) vary in the beating heart, we evaluated contact force-time integral (FTI) as a predictor of lesion size at constant RF power in a contractile bench model simulating the beating heart.

METHODS AND RESULTS

An open-tip irrigated catheter was attached to a movable mount incorporating a dynamic force sensor allowing closed loop control to achieve desired force variations between the catheter tip and bovine skeletal muscle placed on a ground plate. RF energy (20 and 40 W for 60 seconds, 17 cc/min irrigation) was delivered during (1) constant contact (C) at 20 g, (2) variable contact (V) with a 20 g peak and 10 g nadir, and (3) intermittent contact (I) with a 20 g peak and 0 g nadir with loss of contact. V and I protocols were performed at 50 and 100 catheter movements/min and 2 systole:diastole time ratios (50:50 and 30:70). The area under the CF curve was calculated as the FTI. Measured FTI was highest in C, intermediate during V and lowest during I and correlated linearly with lesion volume (P < 0.0001 for 20 and 40 W). Lesion volume was highest in group C, intermediate in V and lowest in group I (P < 0.05 for C vs V, V vs I, and C vs I).

CONCLUSIONS

Lesion size correlates linearly with measured contact FTI. Constant contact produces the largest and intermittent contact the smallest lesions despite constant RF power and identical peak contact forces.

摘要

简介

电极组织接触、射频 (RF) 功率和持续时间是 RF 损伤大小的主要决定因素。由于在跳动的心脏中接触力 (CF) 会发生变化,因此我们在模拟跳动心脏的收缩台模型中评估了接触力-时间积分 (FTI),作为在恒定 RF 功率下预测损伤大小的指标。

方法和结果

将开口尖端灌洗导管连接到一个可移动的支架上,该支架包含一个动态力传感器,允许闭环控制以在导管尖端和放置在接地板上的牛骨骼肌之间实现所需的力变化。在以下情况下输送 RF 能量(20 和 40 W 持续 60 秒,17 cc/min 灌洗):(1) 在 20 g 时进行恒定接触 (C),(2) 以 20 g 峰值和 10 g 谷值进行可变接触 (V),以及 (3) 以 20 g 峰值和 0 g 谷值进行间歇接触 (I) 并失去接触。V 和 I 方案以 50 和 100 个导管运动/分钟以及 2 个收缩:舒张时间比(50:50 和 30:70)进行。作为 FTI,计算 CF 曲线下的面积。测量的 FTI 在 C 中最高,在 V 中居中,在 I 中最低,并且与损伤体积呈线性相关(20 和 40 W 时均 P <0.0001)。损伤体积在 C 组中最高,在 V 组中居中,在 I 组中最低(C 与 V、V 与 I 和 C 与 I 相比,P <0.05)。

结论

损伤大小与测量的接触 FTI 呈线性相关。尽管 RF 功率和相同的峰值接触力保持不变,但恒定接触会产生最大的损伤,而间歇接触会产生最小的损伤。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验