Health Phys. 2010 May;98 Suppl 2:S63-8. doi: 10.1097/HP.0b013e3181d27376.
Measurement of 222Rn is an important research concern because of the high-energy alpha emission of its 218Po and 214Po progeny. The LR-115 passive detector is the basis of a solid-state nuclear track detector (SSNTD) whose use for radon measurements has been well established. The usage of the SSNTD technique requires calibration by comparison. The aim of this work is to evaluate the efficiency calibration parameters for the radon measurement system that three steps: (1) irradiation conditions, (2) etching conditions, and (3) track counting. The following parameters were evaluated: breakdown and counting voltages for the spark counter, detector thickness, etching time, temperature and NaOH concentration as etching conditions for the LR-115 detector. Calibrating factors of 0.035 and 0.028 tracks cm-2 Bq-1 m3 d-1 for the open- and closed-chamber radon monitors, respectively, were determined considering the best efficiency calibration parameter values.
测量 222Rn 是一个重要的研究课题,因为其 218Po 和 214Po 子体具有高能α发射。LR-115 型被动探测器是固态核径迹探测器(SSNTD)的基础,其在氡测量中的应用已得到充分证实。SSNTD 技术的使用需要通过比较进行校准。本工作的目的是评估氡测量系统的效率校准参数,分为三个步骤:(1)辐照条件,(2)蚀刻条件,和(3)径迹计数。评估了以下参数:火花计数器的击穿和计数电压、探测器厚度、蚀刻时间、温度和 NaOH 浓度作为 LR-115 探测器的蚀刻条件。考虑到最佳效率校准参数值,分别确定了开放室和密闭室氡监测器的校准因子为 0.035 和 0.028 条/cm2 Bq-1 m3 d-1。