LGC Ltd., Queen's Rd, Teddington TW11 0LY, UK.
Rapid Commun Mass Spectrom. 2010 May 30;24(10):1392-6. doi: 10.1002/rcm.4539.
High-field asymmetric waveform ion mobility spectrometry (FAIMS) is an ion-filtering technique recently adapted for use with liquid chromatography/mass spectrometry (LC/MS) to remove interferences during analysis of complex matrices. This is the first systematic study of a series of singly charged tetraalkylammonium ions by FAIMS-MS. The compensation voltage (CV) is the DC offset of the waveform which permits the ion to emerge from FAIMS and it was determined for each member of the series under various conditions. The electrospray ionization conditions explored included spray voltage, vaporizer temperature, and sheath and auxiliary gas pressure. The FAIMS conditions explored included carrier gas flow rate, electrode temperature and composition of the carrier gas. Optimum desolvation was achieved using sufficient carrier gas (flow rate > or = 2 L/min) to ensure stable response. Low-mass ions (m/z 100-200) are more susceptible to changes in electrode temperature and gas composition than high mass ions (m/z 200-700). As a result of this study, ions are reliably analyzed using standard FAIMS conditions (dispersion voltage -5000 V, carrier gas flow rate 3 L/min, 50% helium/50%nitrogen, inner electrode temperature 70 degrees C and outer electrode temperature 90 degrees C). Variation of FAIMS conditions may be of great use for the separation of very low mass tetraalkylammonium (TAA) ions from other TAA ions. The FAIMS conditions do not appear to have a major effect on higher mass ions.
高场非对称波形离子淌度谱(FAIMS)是一种最近被应用于液相色谱/质谱(LC/MS)的离子过滤技术,用于在分析复杂基质时去除干扰。这是首次对一系列单价四烷基铵离子进行 FAIMS-MS 的系统研究。补偿电压(CV)是使离子从 FAIMS 中逸出的波形直流偏移量,在各种条件下确定了该系列中每个成员的补偿电压。所探索的电喷雾电离条件包括喷雾电压、汽化器温度以及鞘气和辅助气压力。所探索的 FAIMS 条件包括载气流速、电极温度以及载气的组成。通过使用足够的载气(流速>或=2 L/min)来实现最佳的去溶剂化,从而确保稳定的响应。低质量离子(m/z 100-200)比高质量离子(m/z 200-700)更容易受到电极温度和气体组成变化的影响。由于这项研究,使用标准 FAIMS 条件(色散电压-5000 V,载气流速 3 L/min,50%氦气/50%氮气,内电极温度 70°C 和外电极温度 90°C)可以可靠地分析离子。FAIMS 条件的变化可能对非常低质量的四烷基铵(TAA)离子与其他 TAA 离子的分离非常有用。FAIMS 条件似乎对高质量离子没有重大影响。