To assess in vitro the three-dimensional mean velocity field and the extent and degree of turbulence intensity (TI) in different prosthetic heart valves using a generalization of phase-contrast MRI (PC-MRI).
Four 27-mm aortic valves (Björk-Shiley Monostrut tilting-disc, St. Jude Medical Standard bileaflet, Medtronic Mosaic stented and Freestyle stentless porcine valve) were tested under steady inflow conditions in a Plexiglas phantom. Three-dimensional PC-MRI data were acquired to measure the mean velocity field and the turbulent kinetic energy (TKE), a direction-independent measure of TI.
Velocity and TI estimates could be obtained up- and downstream of the valves, except where metallic structure in the valves caused signal void. Distinct differences in the location, extent, and peak values of velocity and TI were observed between the valves tested. The maximum values of TKE varied between the different valves: tilting disc, 100 J/m(3); bileaflet, 115 J/m(3); stented, 200 J/m(3); stentless, 145 J/m(3).
The TI downstream from a prosthetic heart valve is dependent on the specific valve design. Generalized PC-MRI can be used to quantify velocity and TI downstream from prosthetic heart valves, which may allow assessment of these aspects of prosthetic valvular function in postoperative patients.
使用相位对比 MRI(PC-MRI)的推广方法,评估不同人工心脏瓣膜的三维平均速度场以及湍流强度(TI)的程度和范围。
在有机玻璃模型中,对 4 个 27mm 的主动脉瓣(Björk-Shiley Monostrut 倾斜碟瓣、St. Jude Medical Standard 双叶瓣、Medtronic Mosaic 支架瓣和 Freestyle 无支架猪瓣)进行稳态流入条件下的测试。采集三维 PC-MRI 数据以测量平均速度场和湍流动能(TKE),这是 TI 的一种无方向依赖性的度量。
除了瓣膜中的金属结构导致信号缺失的情况外,在瓣膜的上下游都可以获得速度和 TI 的估计值。测试的瓣膜之间观察到速度和 TI 的位置、程度和峰值存在明显差异。不同瓣膜的 TKE 值最大值有所不同:倾斜碟瓣 100 J/m³;双叶瓣 115 J/m³;支架瓣 200 J/m³;无支架瓣 145 J/m³。
人工心脏瓣膜下游的 TI 取决于特定的瓣膜设计。广义 PC-MRI 可用于量化人工心脏瓣膜下游的速度和 TI,这可能允许评估术后患者这些人工瓣膜功能方面的情况。