Institut National de la Sante et de la Recherche Medicale (Inserm), U556 Lyon, France.
IEEE Trans Ultrason Ferroelectr Freq Control. 2010 May;57(5):1086-95. doi: 10.1109/TUFFC.2010.1520.
Interstitial ultrasound applicators can be a minimally invasive alternative for treating targets that are unresectable or are inaccessible by extracorporeal methods. Dual-mode transducers for ultrasound imaging and therapy were developed to address the constraints of a miniaturized applicator and real-time treatment monitoring. We propose an original treatment strategy that combines ultrasound imaging and therapy using a dual-mode transducer rotating at 8 revolutions per second. Real-time B-mode imaging was interrupted to emit high-intensity ultrasound over a selected therapy aperture. A full 360 degrees image was taken every 8th rotation to image the therapy aperture. Numerical simulations were performed to study the effect of rotation on tissue heating, and to study the effect of the treatment sequence on transducer temperature. With the time-averaged transducer surface intensity held at 12 W/cm(2) to maintain transducer temperature below 66 degrees C, higher field intensities and deeper lesions were produced by narrower therapy apertures. A prototype system was built and tested using in vitro samples of porcine liver. Lesions up to 8 mm were produced using a time-averaged transducer surface intensity of 12 W/cm(2) applied for a period of 240 s over a therapy aperture of 40 degrees. Apparent strain imaging of the therapy aperture improved the contrast between treated and spared tissues, which could not be differentiated on B-mode images. With appropriate limits on the transducer output, real-time imaging and deep thermal ablation are feasible and sustainable using a rotating dual-mode transducer.
间质超声治疗探头可以作为一种微创选择,用于治疗无法切除或无法通过体外方法治疗的目标。双模式超声成像和治疗探头旨在解决小型化探头和实时治疗监测的限制。我们提出了一种新的治疗策略,即使用每秒旋转 8 转的双模式探头进行超声成像和治疗的联合。实时 B 型超声成像会被中断,以在选定的治疗孔径上发射高强度超声。每隔 8 次旋转采集一次完整的 360 度图像,以对治疗孔径进行成像。进行了数值模拟,以研究旋转对组织加热的影响,以及治疗顺序对探头温度的影响。为了将探头温度保持在 66°C 以下,将时间平均探头表面强度保持在 12 W/cm²,从而产生更高的场强和更深的病变。建立了一个原型系统,并使用猪肝的体外样本进行了测试。使用时间平均探头表面强度为 12 W/cm² 的治疗孔径为 40°,施加 240 s,产生了最大 8mm 的病变。治疗孔径的表观应变成像提高了处理组织和未处理组织之间的对比度,B 型图像无法区分这两种组织。在适当限制探头输出的情况下,使用旋转双模式探头可以实现实时成像和深层热消融。