Department of Radiology, Korea University Ansan Hospital, Korea University College of Medicine, 516 Gojan-1-dong, Danwon-gu, Ansan-si, Gyeonggi-do, Ansan, Korea.
Acad Radiol. 2010 Jun;17(6):744-51. doi: 10.1016/j.acra.2010.02.008.
The use of breast computed tomography (CT) has been limited by radiation hazard and image quality. The objective of this study was to compare radiation doses and image quality for different 64-channel multidetector row CT (MDCT) parameters, and to potentially provide optimal CT parameters for breast imaging.
For assessment of radiation doses, CT dose index (CTDI(100)) values were obtained at various x-ray tube voltages (80, 120, 140 kVp) and currents (30, 50, 100, 150, 200 mAs) using a standard CT body dose phantom. To evaluate image quality, four fresh mastectomy specimens were scanned and three radiologists graded images for overall image quality, glandular tissue-fat conspicuity, and Cooper's ligament sharpness. Statistically, linear regression analyses and multiple comparisons were used for investigation of the relationship between radiation dose, image qualities, and CT parameters.
CTDI(100) values of < or =6 mGy were obtained at 80 kVp and any mAs, 120 kVp and 30 or 50 mAs, and 140 kVp and 30 or 50 mAs. Image quality at 80 kVp and 200 mAs, 120 kV and 50, 100, 150, or 200 mAs, and 140 kVp and all mAs values tested were significantly superior to those at 80 kVp and 30, 50, 100, or 150 mAs and 120 kV and 30 mAs (P < .05).
Based on our results, 80 kVp and 200 mAs, 120 kVp and 50 mAs, 140 kVp and 30 mAs, or 140 kVp and 50 mAs can be used for breast MDCT scanning to reduce radiation dose and preserve image quality and 140 kVp at 30 mAs is the optimal setting.
由于辐射危害和图像质量的限制,乳腺计算机断层扫描(CT)的应用受到了限制。本研究的目的是比较不同 64 排多层 CT(MDCT)参数的辐射剂量和图像质量,并为乳腺成像提供潜在的最佳 CT 参数。
为评估辐射剂量,在标准 CT 体模中采用不同的 X 射线管电压(80、120、140 kVp)和电流(30、50、100、150、200 mAs)获取 CT 剂量指数(CTDI(100))值。为了评估图像质量,对 4 例新鲜乳房切除术标本进行了扫描,3 位放射科医生对整体图像质量、腺体组织-脂肪对比和 Cooper 韧带锐利度进行了评分。采用线性回归分析和多重比较对辐射剂量、图像质量和 CT 参数之间的关系进行了统计学研究。
在 80 kVp 和任意 mAs、120 kVp 和 30 或 50 mAs 以及 140 kVp 和 30 或 50 mAs 时,可获得<或=6 mGy 的 CTDI(100)值。80 kVp 和 200 mAs、120 kV 和 50、100、150 或 200 mAs 以及 140 kVp 和所有 mAs 值的图像质量明显优于 80 kVp 和 30、50、100 或 150 mAs 以及 120 kV 和 30 mAs(P<0.05)。
根据我们的结果,80 kVp 和 200 mAs、120 kVp 和 50 mAs、140 kVp 和 30 mAs 或 140 kVp 和 50 mAs 可用于乳腺 MDCT 扫描以降低辐射剂量并保持图像质量,而 140 kVp 时 30 mAs 是最佳设置。