Department of Orthodontics, Dental Clinic, Georg-August-University of Göttingen, Germany.
Eur J Orthod. 2010 Dec;32(6):607-13. doi: 10.1093/ejo/cjq007. Epub 2010 May 12.
The forces delivered by aligners during torquing have still not been investigated. The purpose of this study was to measure the forces delivered to an upper central incisor during torquing with three different materials of the same thickness, and to describe the biomechanical principles of torquing with aligners. Five identical appliances were manufactured from each of three materials, all with a thickness of 1.0 mm (Ideal Clear®, Erkodur®, and Biolon®). An upper central incisor, as part of the measuring device, was torqued in defined steps in the vestibular and palatal directions with the respective appliance in place. For statistical analysis, the resulting forces, Fx (forces acting in the palatal and facial directions) and Fz (intrusive force as a side-effect) at a displacement of ±0.15 and ±0.8 mm from the tooth at the gingival margin were calculated. The mean Fx forces for ±0.15 mm displacement ranged from -1.89 N [standard deviation (SD) 0.48] to 0.11 N (SD 0.1). The mean Fz forces were between -0.97 N (SD 0.57) and -0.07 N (SD 0.22). The highest intrusive forces were measured during palatal displacement of the measuring tooth. An influence of direction of displacement on the levels of force was observed, especially for Fz at the greater displacement of ±0.8 mm. In relation to the intended amount of root movement during torquing, aligners tend to 'lift up' and therefore no effective force couple can be established for further root control. The force delivery properties are also influenced by the material used and the shape of the tooth.
矫正器在扭转过程中产生的力尚未得到研究。本研究的目的是测量扭转过程中三种相同厚度的材料对上中切牙施加的力,并描述用矫正器进行扭转的生物力学原理。从三种材料中各制造了五个相同的矫正器,厚度均为 1.0 毫米(理想透明矫正器、Erkodur 和 Biolon)。在上中切牙作为测量装置的一部分,在颊侧和腭侧以规定的步骤用相应的矫正器进行扭转。为了进行统计分析,在牙冠边缘牙龈处的牙位移为±0.15 和±0.8 毫米时,计算出 Fx(在腭侧和颊侧起作用的力)和 Fz(作为副作用的侵入力)的合力。对于±0.15 毫米的位移,平均 Fx 力范围为-1.89N(标准偏差(SD)0.48)至 0.11N(SD 0.1)。平均 Fz 力在-0.97N(SD 0.57)和-0.07N(SD 0.22)之间。在测量牙的腭向位移时,测量到的侵入力最大。观察到位移方向对力水平的影响,尤其是在更大的位移±0.8 毫米时的 Fz。相对于扭转过程中预期的根移动量,矫正器往往会“抬起”,因此无法为进一步的根控制建立有效的力偶。力传递特性还受到所使用的材料和牙齿形状的影响。