Schneider C J, Engelberts N, van Dijk J D
Department of Radiotherapy, Academisch Medisch Centrum, Amsterdam, The Netherlands.
Phys Med Biol. 1991 Apr;36(4):461-74. doi: 10.1088/0031-9155/36/4/005.
An experimental study is described, leading to the design of a near-field electromagnetic radiation probe. A passive field probe with fibre-optic link has been constructed for use in hyperthermia phantoms, i.e. for measurements in lossy media--saline was used in this study as a simulation of biological tissue. The realization of a closed electric circuit on the probe proved to be a major influence on the sensitivity and dynamic range of the probe. The experiments also independently confirmed the dipole length, frequency variations between 5 and 110 MHz and the temperature of the probe as important parameters.
本文描述了一项实验研究,该研究促成了一种近场电磁辐射探头的设计。已构建了一种带有光纤链路的无源场探头,用于热疗体模,即在有损介质中进行测量——本研究中使用盐水模拟生物组织。事实证明,在探头上实现闭合电路对探头的灵敏度和动态范围有重大影响。实验还独立证实了偶极子长度、5至110兆赫兹之间的频率变化以及探头温度是重要参数。