J Chromatogr A. 2010 Jun 25;1217(26):4439-47. doi: 10.1016/j.chroma.2010.04.051. Epub 2010 Apr 24.
A novel gas chromatography/high resolution mass spectrometry method coupled with multisorbent thermal desorption cartridges has been developed for the determination of volatile neutral polyfluorinated iodine alkanes (PFIs) in airborne samples. It allows, for the first time, simultaneous analysis of four mono-iodized perfluorinated alkanes, three diiodized perfluorinated alkanes and four mono-iodized polyfluorinated telomers in ambient air samples. 3.75 L air sample was passed through a sorbent tube packed with 150 mg of Tenax TA and 200mg of Carbograph 1 TD for analyte adsorption. Important factors during the analysis procedures, such as safe sampling volume, air sampling rate, analyte desorption and transfer strategies, were optimized and good thermal desorption efficiencies were obtained. The method detection limit (MDL) concentration ranged from 0.04 pg/L for 1H,1H,2H,2H-perfluorododecyl iodide to 1.2 pg/L for perfluorohexyl iodide, and instrument response of a seven-point calibration was linear in the range of 10-1000 pg. Travel spike recoveries ranged from 83% to 116%. Small variabilities of less than 36% were obtained near the MDLs and the differences between triplicates were even smaller (2.1-7.3%) at 200 pg spiked level. The method was successfully applied to analyze ambient air samples collected near a point source, and five PFIs were identified (10.8-85.0 pg/L), with none of the analytes detectable at the background site.
一种新型的气相色谱/高分辨质谱联用多吸附热脱附管方法已经被开发出来,用于测定空气中挥发性中性多氟碘烷(PFIs)。它首次允许同时分析环境空气中的四种单碘代全氟烷烃、三种二碘代全氟烷烃和四种单碘代多氟端基聚合物。3.75 L 空气样品通过装有 150mg Tenax TA 和 200mg Carbograph 1 TD 的吸附管进行分析物吸附。在分析过程中,重要因素,如安全采样体积、空气采样速率、分析物解吸和转移策略等,都进行了优化,并获得了良好的热脱附效率。方法检测限(MDL)浓度范围为 0.04pg/L(1H,1H,2H,2H-全氟十二基碘)至 1.2pg/L(全氟己基碘),七点校准的仪器响应在线性范围为 10-1000pg。旅行尖峰回收率在 83%至 116%之间。在接近 MDL 时,可获得小于 36%的小变异性,而在 200pg 加标水平时,三重空白之间的差异甚至更小(2.1-7.3%)。该方法成功地应用于分析点源附近采集的环境空气样品,并鉴定出五种 PFIs(10.8-85.0pg/L),在背景点没有检测到任何分析物。