Suppr超能文献

平板三维成像引导下微创胸腰椎椎弓根螺钉置入:一项可行性研究

Navigated minimally invasive thoracolumbar pedicle screw placement with flat panel 3-D imaging. A feasibility study.

作者信息

Citak Musa, Stubig Timo, Kendoff Daniel, Citak Mustafa, O'Loughlin Padhraig F, Hüfner Tobias, Krettek Christian

机构信息

Trauma Department, Hannover Medical School, Hannover, Germany.

出版信息

Technol Health Care. 2010;18(2):101-10. doi: 10.3233/THC-2010-0572.

Abstract

STUDY DESIGN

The current study involves a cadaveric specimen with navigated pedicle screw placement using conventional reference markers and isocentric 3D fluoroscopy and also minimally invasive reference marker use with a flat panel 3D scanner.

OBJECTIVE

To test the feasibility of a novel minimally invasive reference marker system for navigated pedicle screw placement in combination the use of a new imaging modality i.e. flat panel 3D imaging.

SUMMARY OF BACKGROUND DATA

A major limiting factor of navigated pedicle screw placement is the requirement for intraosseous fixation of reference markers. This usually necessitates an open approach to the spinous process. To address this issue, the current authors have developed a minimally-invasive fixation device for spinal reference marker fixation.

METHODS

A fresh-frozen cadaver with no history of spinal injury or deformity was positioned prone on a radiolucent table. L3 and L4 vertebrae were randomly selected for conventional pedicle screw insertion while T5 and T6 were selected for the percutaneous technique. A flat detector 3D C-arm (Ziehm vision FD 3D; Ziehm, Nuremberg, Germany) was used to evaluate the position of the pedicle screws at the vertebral levels targeted in the study.

RESULTS

All screws placed within the lumbar spine involved conventional reference markers. The average depth deviation was 0.73 mm and the average axis deviation was 1.67 mm. Within the thoracic spine, the minimally-invasive marker fixation devices were used. The average depth deviation was 0.85 mm and the average axis deviation was 1.75 mm. In both cases, the plan and navigation were performed satisfactorily. The Y-jaw clamp for minimally-invasive reference fixation seemed to provide stable and robust fixation of the markers, requiring only two small incisions.

CONCLUSIONS

The minimally invasive reference marker system produced results which were comparable to the conventional intra-osseous markers while the flat detector-based navigation was shown to be easier to use and faster than isocentric Iso-3D technology.

摘要

研究设计

本研究涉及一个尸体标本,使用传统参考标记和等中心三维荧光透视进行导航椎弓根螺钉置入,同时使用平板三维扫描仪进行微创参考标记的应用。

目的

测试一种新型微创参考标记系统在结合使用新型成像方式即平板三维成像时用于导航椎弓根螺钉置入的可行性。

背景数据总结

导航椎弓根螺钉置入的一个主要限制因素是参考标记的骨内固定要求。这通常需要对棘突采用开放入路。为解决这一问题,作者们开发了一种用于脊柱参考标记固定的微创固定装置。

方法

将一具无脊柱损伤或畸形病史的新鲜冷冻尸体俯卧于可透射线的手术台上。随机选择L3和L4椎体进行传统椎弓根螺钉置入,而选择T5和T6椎体采用经皮技术。使用平板探测器三维C型臂(Ziehm vision FD 3D;德国纽伦堡的Ziehm公司)评估研究中目标椎体水平的椎弓根螺钉位置。

结果

腰椎置入的所有螺钉均采用传统参考标记。平均深度偏差为0.73毫米,平均轴线偏差为1.67毫米。在胸椎,使用了微创标记固定装置。平均深度偏差为0.85毫米,平均轴线偏差为1.75毫米。在这两种情况下,计划和导航均执行得令人满意。用于微创参考固定的Y形夹似乎能为标记提供稳定而牢固的固定,仅需两个小切口。

结论

微创参考标记系统产生的结果与传统骨内标记相当,同时基于平板探测器的导航被证明比等中心Iso-3D技术更易于使用且更快。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验