Department of Electrical Engineering, Stanford University, Stanford, CA 94305, USA.
Physiol Meas. 2010 Jul;31(7):N51-9. doi: 10.1088/0967-3334/31/7/N02. Epub 2010 Jun 4.
Ballistocardiography is a non-invasive technique for evaluating cardiovascular health. This note presents an ac-bridge amplifier for low-noise ballistocardiogram (BCG) recording from a modified weighing scale. The strain gauges in a commercial scale were excited by an ac source-square or sine wave-and the differential output voltage resulting from the BCG was amplified and demodulated synchronously with the excitation waveform. A standard BCG amplifier, with a simple dc-bridge excitation, was also built and the performance was compared to both the square- and sine-wave excited ac-bridge amplifiers. The total input-referred voltage noise (rms) integrated over the relevant BCG bandwidth of 0.3-10 Hz was found to be 30 nV (square wave source) or 25 nV (sine-wave source) for the ac-bridge amplifier and 52 nV for the standard amplifier: an improvement of 4.8 dB or 6 dB, respectively. These correspond to input-referred force noise (rms) values of 5 mN, 4 mN and 8.3 mN. The improvement in SNR was also observed in recorded waveforms from a seated subject whose BCG signal was measured with both dc- and ac-bridge circuits.
心动描记术是一种评估心血管健康的非侵入性技术。本说明介绍了一种用于从改装的称重秤记录低噪声心动描记图(BCG)的交流桥放大器。商用秤中的应变计由交流电源方波或正弦波激励,BCG 产生的差分输出电压与激励波形同步放大和解调。还构建了一个具有简单直流桥激励的标准 BCG 放大器,并将其性能与方波和正弦波激励的交流桥放大器进行了比较。在 0.3-10 Hz 的相关 BCG 带宽上,总输入参考电压噪声(rms)被发现为交流桥放大器的 30 nV(方波源)或 25 nV(正弦波源),而标准放大器为 52 nV:分别提高了 4.8 dB 或 6 dB。这对应于输入参考力噪声(rms)值为 5 mN、4 mN 和 8.3 mN。在使用直流和交流桥电路测量坐姿受试者的 BCG 信号时,也观察到 SNR 的改善。