This study aims to investigate the mechanics of the intra-aortic balloon (IAB) under different aortic pressure (P(ao)) and inclination (0-75 degrees). Pressure and flow were measured in an artificial aorta during IAB pumping with a frequency of 1:3. Volume displaced toward the "coronary arteries" during inflation (V(prox)) and "intra-aortic" pressure reduction during deflation (P(r)) were derived. IAB duration of inflation and deflation was determined with a high-speed camera visualization. When the aorta was horizontal, P(ao) raised from 45 mm Hg to 115 mm Hg, V(prox) reduced by 18% (25.0 +/- 1.0 mL vs. 30.4 +/- 1.9 mL) and P(r) increased by 117% (106.4 +/- 0.3 mm Hg vs. 48.9 +/- 0.6 mm Hg). When the aorta was inclined, at low P(ao) of 45 mm Hg, V(prox) was reduced by 30% from 0 degrees to 45 degrees (19.8 +/- 2.3 mL vs. 28.3 +/- 1.7 mL) and P(r) was reduced by 66% (16.5 +/- 0.1 mm Hg vs. 48.9 +/- 0.6 mm Hg). However, at high P(ao) of 115 mm Hg, V(prox) remained unchanged with increasing angle (20.0 +/- 1.0 mL) and P(r) was reduced by 24% (80.6 +/- 0.8 mm Hg vs. 106.4 +/- 0.3 mm Hg). Increasing P(ao) increased duration of inflation. At low P(ao), increasing angle resulted in increasing duration of inflation, but at high P(ao), increasing angle had the opposite effect. Duration of deflation generally decreased with P(ao) and increased with increasing angle. The IAB pump is affected by both P(ao) and angle, indicating that non-normotensive patients or patients in the semi-recumbent position might not receive the full benefits of IAB counterpulsation.
本研究旨在探讨不同主动脉压(P(ao))和倾斜角度(0-75 度)下主动脉内球囊(IAB)的力学特性。在 IAB 以 1:3 的频率进行泵血时,在人工主动脉中测量压力和流量。在充气期间朝向“冠状动脉”的位移体积(V(prox))和放气期间的“主动脉内”压力降低(P(r))。通过高速摄像机可视化确定 IAB 的充气和放气持续时间。当主动脉处于水平位置时,P(ao)从 45mmHg 升高到 115mmHg,V(prox)降低 18%(25.0 +/- 1.0mL 对 30.4 +/- 1.9mL),P(r)升高 117%(106.4 +/- 0.3mmHg 对 48.9 +/- 0.6mmHg)。当主动脉倾斜时,在低 P(ao)为 45mmHg 时,V(prox)从 0 度到 45 度降低 30%(19.8 +/- 2.3mL 对 28.3 +/- 1.7mL),P(r)降低 66%(16.5 +/- 0.1mmHg 对 48.9 +/- 0.6mmHg)。然而,在高 P(ao)为 115mmHg 时,随着角度的增加,V(prox)保持不变(20.0 +/- 1.0mL),P(r)降低 24%(80.6 +/- 0.8mmHg 对 106.4 +/- 0.3mmHg)。增加 P(ao)会增加充气持续时间。在低 P(ao)下,增加角度会导致充气持续时间增加,但在高 P(ao)下,增加角度会产生相反的效果。放气持续时间通常随 P(ao)的增加而减少,随角度的增加而增加。IAB 泵受到 P(ao)和角度的双重影响,这表明非正常血压患者或半卧位患者可能无法获得 IAB 反搏的全部益处。