Eur J Mass Spectrom (Chichester). 2010;16(3):397-406. doi: 10.1255/ejms.1076.
The mass resolution of a time-of-flight (ToF) mass spectrometer is directly proportional to its total flight path length. We have developed multi-turn ToF mass spectrometers, where ions are stored in a fixed orbit within electrostatic sectors and allowed to propagate the said orbit numerous times. With each successive orbit, the flight path is correspondingly increasing. The first multi-turn ToF mass spectrometer, the MULTUM Linear plus, was developed for cometary exploration. The spectrometer consists of four cylindrical electrostatic sectors and 28 electrostatic quadrupole lenses. The size of the analyzer is 40 cm square. Mass resolution is demonstrated to increase according to the number of ion cycles. A mass resolution of greater than 350,000 was achieved after 501.5 cycles. Another multi-turn ToF mass spectrometer, the MULTUM II, which consists of only four toroidal electrostatic sectors, was also developed in an effort to reduce the number of quadrupole lenses. We are developing various types of mass spectrometer based on the MULTUM II technology, a ToF/ToF mass spectrometer "MULTUM- TOF/TOF", a stigmatic imaging mass spectrometer "MULTUM-IMG" and miniature mass spectrometers of high mass resolving power, the "MULTUM-S" series.
飞行时间(ToF)质谱仪的质量分辨率与其总飞行路径长度成正比。我们已经开发出多圈ToF质谱仪,其中离子被存储在静电扇区内的固定轨道中,并允许多次传播该轨道。随着每一次连续的轨道运行,飞行路径相应增加。第一台多圈ToF质谱仪“MULTUM Linear plus”是为彗星探测而开发的。该光谱仪由四个圆柱形静电扇区和28个静电四极透镜组成。分析仪的尺寸为40平方厘米见方。质量分辨率被证明会随着离子循环次数的增加而提高。在501.5次循环后实现了大于350,000的质量分辨率。另一台多圈ToF质谱仪“MULTUM II”也已开发出来,它仅由四个环形静电扇区组成,旨在减少四极透镜的数量。我们正在基于“MULTUM II”技术开发各种类型的质谱仪,一种ToF/ToF质谱仪“MULTUM-TOF/TOF”、一种像散成像质谱仪“MULTUM-IMG”以及具有高质量分辨能力的微型质谱仪“MULTUM-S”系列。