评估早产儿在产房行面罩通气时的潮气量和气体泄漏。

Assessment of tidal volume and gas leak during mask ventilation of preterm infants in the delivery room.

机构信息

Neonatal Services, Department of Newborn Research, The Royal Women's Hospital, Melbourne, Australia.

出版信息

Arch Dis Child Fetal Neonatal Ed. 2010 Nov;95(6):F393-7. doi: 10.1136/adc.2009.174003. Epub 2010 Jun 14.

Abstract

AIM

The aim was to compare resuscitators' estimates of tidal volume (V(T)) and face mask leak with measured values during positive pressure ventilation (PPV) of newborn infants in the delivery room.

PATIENTS AND METHODS

The authors measured inflating pressures and V(T) delivered using a respiratory function monitor, and calculated face mask leak. After 60 s of PPV, resuscitators were asked to estimate V(T) and face mask leak. These estimates were compared with measurements taken during the previous 30 s.

RESULTS

The authors studied 20 infants who received a mean (SD) of 21 (6) inflations during the 30 s. The median (IQR) expired tidal volume (V(Te)) delivered was 8.7 ml/kg (5.3-11.3). V(Te) varied widely during each resuscitation and between resuscitators. Five resuscitators could not estimate V(Te), one overestimated and 14 underestimated their median delivered V(Te). The median (IQR) face mask leak was 29% (16-63%). Leak also varied widely during each resuscitation and between resuscitators. One resuscitator could not estimate mask leak, four overestimated leak and 15 underestimated leak.

CONCLUSION

During face mask ventilation in the delivery room, V(T) and face mask leak were large and variable. The resuscitators were unable to accurately assess their face mask leak or delivered V(T).

摘要

目的

旨在比较复苏器在产房对新生儿进行正压通气(PPV)时对潮气量(V(T))和面罩泄漏的估计与实际测量值。

患者和方法

作者使用呼吸功能监测仪测量充气压力和输送的 V(T),并计算面罩泄漏。在进行 60 秒的 PPV 后,要求复苏器估计 V(T)和面罩泄漏。将这些估计值与前 30 秒的测量值进行比较。

结果

作者研究了 20 名接受平均(SD)21(6)次通气的婴儿,在 30 秒内。输送的呼气潮气量(V(Te))中位数(IQR)为 8.7 ml/kg(5.3-11.3)。每次复苏和复苏器之间的 V(Te)差异很大。五名复苏器无法估计 V(Te),一名估计过高,14 名估计过低。面罩泄漏中位数(IQR)为 29%(16-63%)。泄漏在每次复苏和复苏器之间也差异很大。一名复苏器无法估计面罩泄漏,四名估计过高,15 名估计过低。

结论

在产房进行面罩通气时,V(T)和面罩泄漏量较大且变化较大。复苏器无法准确评估其面罩泄漏或输送的 V(T)。