Jones D L, Irish W D, Klein G J
Department of Medicine, University of Western Ontario, London, Canada.
Circ Res. 1991 Jul;69(1):45-51. doi: 10.1161/01.res.69.1.45.
When an automatic defibrillator is implanted, it is essential to determine the efficacy of the defibrillating system accurately, while balancing the need to keep the number of fibrillation episodes to a minimum. Two methods have evolved to assess defibrillation efficacy: 1) the "defibrillation threshold," which requires few ventricular fibrillation episodes, and 2) the "dose-response curve," which requires many ventricular fibrillation episodes and relates percent success to energy. The purpose of this study was to compare these two methods directly. Twenty open-chest anesthetized pigs had triplicate defibrillation threshold determinations. To produce a dose-response curve, six shocks then were delivered at 0.5, 0.75, 1.0, 1.25, 1.5, and 2.0 times the mean defibrillation threshold, in a balanced randomized order, during separate episodes of ventricular fibrillation. The data were fitted by logistic regression, conversions of the logistic regression, and a saturable exponential and nonsaturable growth exponential. A comparison was made of the mean defibrillation threshold and the 50% point on the dose-response curve (ED50) for each model, for each animal. In addition, the reliability of each measure was assessed by comparing the coefficients of variation. There was no statistical difference between the group defibrillation threshold (6.6 +/- 0.5 J) and group ED50 values (ED50 range of the models, 5.7 +/- 1.9 to 7.0 +/- 0.9 J). However, the variability about the defibrillation threshold was less than that of the ED50 values for all mathematical models except the true logistic equation, which was virtually the same.(ABSTRACT TRUNCATED AT 250 WORDS)
植入自动除颤器时,在尽量减少颤动发作次数的同时准确测定除颤系统的疗效至关重要。已发展出两种评估除颤疗效的方法:1)“除颤阈值”,此方法所需的室颤发作次数较少;2)“剂量-反应曲线”,此方法需要多次室颤发作,并将成功百分比与能量相关联。本研究的目的是直接比较这两种方法。20只开胸麻醉猪进行了三次除颤阈值测定。为绘制剂量-反应曲线,在单独的室颤发作期间,以平衡随机顺序,分别给予平均除颤阈值0.5、0.75、1.0、1.25、1.5和2.0倍的六次电击。数据通过逻辑回归、逻辑回归转换以及饱和指数和非饱和增长指数进行拟合。对每个动物的每个模型,比较了平均除颤阈值和剂量-反应曲线上的50%点(ED50)。此外,通过比较变异系数评估了每种测量方法的可靠性。组除颤阈值(6.6±0.5J)与组ED50值(模型的ED50范围为5.7±1.9至7.0±0.9J)之间无统计学差异。然而,除了实际相同的真实逻辑方程外,所有数学模型中除颤阈值的变异性均小于ED50值的变异性。(摘要截断于250字)