Int J Comput Assist Radiol Surg. 2011 Mar;6(2):153-61. doi: 10.1007/s11548-010-0495-7. Epub 2010 Jun 24.
Non-invasive imaging assessment of cardiac function is important in cardiovascular disease diagnosis, especially for evaluation of local cardiac motion. Tagged cardiac MRI has been developed for this purpose, but evaluation of the results requires quantification and automation.
Two methods utilizing active contour modeling for wall motion extraction based on tagged cardiac MRI scans were evaluated based on properties of tracking methods in the image domain and frequency domain. Three criteria were used: accuracy, inter-subject and intra-subject sensitivity. The tracking results were evaluated by a medical expert. The evaluation methodology and its possible generalization to other diagnostic methods were considered.
Image domain and frequency domain analysis of tagged cardiac MRI data sets were evaluated demonstrating that the image domain method provides better results. The image domain method method is much more resistant to changes in the data, this time, due to a different subject being scanned. The frequency domain approach is not suitable for clinical applications, as the global error is significantly increased (more than 20%).
The image domain method was found most effective, and it can generate a set of clearly identified parameters. The evaluation approach can be an interesting alternative to classical psychovisual studies which are time-consuming and often fastidious for clinicians.