健康绝经前和绝经后女性桡骨和胫骨外周定量 CT 测量的可重复性。

Reproducibility of peripheral quantitative computed tomography measurements at the radius and tibia in healthy pre- and postmenopausal women.

机构信息

Department of Nuclear Medicine, Hamilton Health Sciences, McMaster University, Hamilton, Ontario, Canada.

出版信息

Can Assoc Radiol J. 2011 Aug;62(3):183-9. doi: 10.1016/j.carj.2010.04.011. Epub 2010 Jun 23.

Abstract

PURPOSE

The objectives of this study were to utilise the XCT-2000 pQCT scanner to determine the mean values and the reproducibility of in vivo total, trabecular, and cortical volumetric bone measurements at distal and diaphyseal sites of the radius and the tibia, as well as calf muscle and subcutaneous fat areas, in healthy pre- and postmenopausal women.

METHODS

Twenty-nine women (14 premenopausal and 15 postmenopausal) were recruited to participate in this study. Distal and diaphyseal sites of the radius (at 4% and 20% of the length of the radius) and tibia (at 4%, 38%, and 66% of the length of the tibia) were examined.

RESULTS

The root mean square coefficient of variation for measurements at the distal tibia gave the most favorable reproducibility values for total (1.5%) and trabecular (1.6%) density, whereas the diaphyseal tibia showed the most favorable reproducibility value for cortical density (0.3%). The root mean square coefficients of variation for measurements of muscle and fat cross-sectional areas at the calf were 0.6% and 0.7%, respectively. At the distal tibia, the mean values for total (P < .05) and trabecular (P < .01) density were significantly lower in postmenopausal women than in premenopausal women.

CONCLUSIONS

The data presented here indicate that XCT-2000 pQCT scans at the tibia provide highly reproducible measurements of total, cortical, and trabecular bone as well as muscle and fat cross-sectional areas. Furthermore, significant differences in volumetric bone measurements between healthy pre- and postmenopausal women were evident only at the distal tibia, suggesting that this site warrants further study.

摘要

目的

本研究旨在利用 XCT-2000 pQCT 扫描仪确定健康绝经前和绝经后妇女桡骨远端和骨干以及胫骨远端和骨干、小腿肌肉和皮下脂肪区域的总、小梁和皮质体积骨测量的平均值和可重复性。

方法

招募了 29 名女性(14 名绝经前和 15 名绝经后)参与本研究。检查了桡骨(桡骨长度的 4%和 20%处)和胫骨(胫骨长度的 4%、38%和 66%处)的远端和骨干部位。

结果

胫骨远端的总(1.5%)和小梁(1.6%)密度的测量值的均方根变异系数给出了最有利的可重复性值,而骨干胫骨的皮质密度(0.3%)具有最有利的可重复性值。小腿肌肉和脂肪横截面积的均方根变异系数分别为 0.6%和 0.7%。在胫骨远端,绝经后妇女的总(P<.05)和小梁(P<.01)密度平均值明显低于绝经前妇女。

结论

这里提供的数据表明,胫骨 XCT-2000 pQCT 扫描可提供总、皮质和小梁骨以及肌肉和脂肪横截面积的高度可重复测量。此外,健康绝经前和绝经后妇女之间体积骨测量的显着差异仅在胫骨远端明显,这表明该部位值得进一步研究。