J Ultrasound Med. 2010 Jul;29(7):1075-81. doi: 10.7863/jum.2010.29.7.1075.
The goal of this work was to evaluate a possible improvement in ultrasound coverage for a dual-modality breast imaging system in the mammographic geometry.
A pilot study was performed to evaluate use of a rubber dam to retain ultrasound gel and improve imaging coverage at the breast periphery on a combined imaging system consisting of an ultrasound scanner and a digital x-ray tomosynthesis unit. Several dams were constructed to encompass the shapes of various sizes of compressed breasts. Visual tracings of the breast-to-paddle contact area and breast periphery were made for 8 breasts to estimate coverage area. Two readers independently reviewed the resulting images and were asked to rate the overall breast image quality.
The percentages of breast in contact with the paddle were greater (P < .01) and the linear dimensions of breast in contact with the paddle were larger (P < .05) with the rubber dam than without it. With the dam, the mean estimated area of the breast in contact with the paddle increased 14%, whereas the mean increase in the fraction of the total breast area in contact with paddle was 30%. The difference was due to the mean total projected area of the breast decreasing 12% as the dam was pressed against it. The image quality of automated ultrasound with the rubber dam was consistently judged to be superior to that without the dam.
This method can enhance the absolute and percentage area of the breast in contact with the paddle, reducing noncontact gaps at the breast periphery. Gently pressing the breast periphery with the dam inserted toward the chest wall improves coverage in automated breast ultrasound scanning.
本研究旨在评估在乳腺摄影几何形状中,双模态乳腺成像系统的超声覆盖范围是否可能得到改善。
进行了一项试点研究,以评估使用橡胶坝来保留超声凝胶并改善组合成像系统(包括超声扫描仪和数字 X 线断层合成单元)在乳房周边的成像覆盖范围。制作了几个坝来涵盖各种大小压缩乳房的形状。对 8 个乳房进行了乳房与桨接触区域和乳房周边的可视追踪,以估计覆盖区域。两位读者独立审查了所得图像,并被要求对整体乳房图像质量进行评分。
与无坝相比,橡胶坝的桨接触乳房百分比更大(P <.01),桨接触乳房的线性尺寸更大(P <.05)。使用橡胶坝时,与桨接触的乳房估计面积平均增加了 14%,而与桨接触的乳房总面积的分数平均增加了 30%。差异是由于当坝压在乳房上时,乳房的总投影面积平均减少了 12%。带有橡胶坝的自动超声的图像质量始终被判断优于没有橡胶坝的图像质量。
这种方法可以增强与桨接触的乳房的绝对和百分比面积,减少乳房周边的非接触间隙。将橡胶坝插入胸壁以轻轻按压乳房周边可以改善自动乳腺超声扫描的覆盖范围。