Evaluation of demineralized bone and bone transplants in vitro and in vivo with cone beam computed tomography imaging.

机构信息

Clinic for Craniomaxillofacial Surgery, University of Mainz, Augustusplatz 2, 55131 Mainz, Germany.

出版信息

Dentomaxillofac Radiol. 2010 Jul;39(5):264-9. doi: 10.1259/dmfr/147454468.

Abstract

OBJECTIVES

The objective of this study was to determine the ability of two flat panel cone beam CT (CBCT) devices to identify demineralized bone and bone transplants in vivo and in vitro.

METHODS

Datasets from patients with autologous bone grafts (n = 9, KaVo 3DeXam (KaVo, Biberach, Germany); n = 38, Accuitomo 40 (Morita, Osaka, Japan)) were retrospectively evaluated. Demineralized and non-demineralized porcine cancellous bone blocks were examined with the two CBCT devices. A SawBone skull (Pacific Research Laboratories, Vashon, WA) was used as a positioning tool for the bone blocks. Descriptive evaluation and image quality assessment were conducted on the KaVo 3DeXam data (voxel size 0.3 mm) using the OsiriX viewer as well as on the Morita Accuitomo data (voxel size 0.25 mm) using proprietary viewer software.

RESULTS

Both in vivo and in vitro, the descriptive analysis of the images of the two devices showed well-visualized bone transplants with clearly defined cancellous bones and well-defined single bone trabeculae in all cross-sections. In vitro, demineralized samples showed lower radiographic opacity but no significant loss of quality compared with fresh bone (P = 0.070). Single cancellous bone trabeculae were significantly better visualized with the Morita 3D Accuitomo device than with the KaVo 3DeXam device (P = 0.038).

CONCLUSION

Both the KaVo 3DeXam and Morita 3D Accuitomo devices produce good-quality images of cancellous bones in in vivo remodelling as well as after in vitro demineralization.

摘要

目的

本研究旨在确定两种平板锥形束 CT(CBCT)设备在体内和体外识别脱矿骨和骨移植物的能力。

方法

回顾性分析了 9 例自体骨移植物患者(KaVo 3DeXam(KaVo,比伯拉赫,德国);n=38,Accuitomo 40(森田,大阪,日本))的数据集。用两种 CBCT 设备检查了脱矿和未脱矿的猪松质骨块。SawBone 颅骨(太平洋研究实验室,瓦雄,WA)用作骨块的定位工具。使用 OsiriX 查看器对 KaVo 3DeXam 数据(体素大小 0.3mm)进行描述性评估和图像质量评估,并使用专有查看器软件对 Morita Accuitomo 数据(体素大小 0.25mm)进行评估。

结果

体内和体外,两种设备图像的描述性分析均显示,在所有横截面上,骨移植物均得到良好显示,松质骨边界清晰,单个骨小梁清晰可见。体外,脱矿样本的射线照相不透明度较低,但与新鲜骨相比,质量无明显损失(P=0.070)。与 KaVo 3DeXam 设备相比,Morita 3D Accuitomo 设备可更好地显示单个松质骨小梁(P=0.038)。

结论

KaVo 3DeXam 和 Morita 3D Accuitomo 两种设备均能产生体内重建和体外脱矿后松质骨的高质量图像。

相似文献

本文引用的文献

1
New challenges in MDCT.多层螺旋CT的新挑战。
Eur Radiol. 2005 Dec;15 Suppl 5:E35-45. doi: 10.1007/s10406-005-0164-z.
3
[Osirix: free and open-source software for medical imagery].[Osirix:用于医学影像的免费开源软件]
Rev Stomatol Chir Maxillofac. 2008 Feb;109(1):53-5. doi: 10.1016/j.stomax.2007.07.007. Epub 2008 Feb 21.
8
Flat-detector computed tomography (FD-CT).平板探测器计算机断层扫描(FD-CT)。
Eur Radiol. 2007 Nov;17(11):2767-79. doi: 10.1007/s00330-007-0651-9. Epub 2007 Jun 23.
9
X-ray scatter data for flat-panel detector CT.平板探测器CT的X射线散射数据。
Phys Med. 2007 Mar;23(1):3-15. doi: 10.1016/j.ejmp.2006.12.004. Epub 2007 Mar 28.
10
Assessment of bone segmentation quality of cone-beam CT versus multislice spiral CT: a pilot study.锥形束CT与多层螺旋CT骨分割质量的评估:一项初步研究。
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006 Aug;102(2):225-34. doi: 10.1016/j.tripleo.2005.10.039. Epub 2006 Apr 21.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验