Prairie Cardiovascular Consultants, Southern Illinois University, 401 East Carpenter Street, Springfield, IL 62702, USA.
Europace. 2010 Nov;12(11):1654-5. doi: 10.1093/europace/euq189. Epub 2010 Jul 1.
We report a case of inadvertent transarterial implantation of dual-chamber pacemaker leads; the ventricular lead positioned across the aortic valve into the left ventricle and the atrial lead curving in the aortic root with the tip positioned into the left circumflex artery, resulting in acute myocardial infarction. The diagnosis was made based on the finding on the chest X-ray, surface ECG, and coronary angiography.
我们报告了一例双腔起搏器导线意外经动脉植入的病例;心室导线穿过主动脉瓣进入左心室,心房导线在主动脉根部弯曲,尖端位于左回旋支动脉内,导致急性心肌梗死。诊断依据为胸部 X 线、体表心电图和冠状动脉造影的发现。