Department of Nephrology, Ruijin Hospital, Shanghai Jiaotong University, School of Medicine, Shanghai, Peoples Republic of China.
Am J Kidney Dis. 2010 Nov;56(5):1006-11. doi: 10.1053/j.ajkd.2010.03.034. Epub 2010 Jul 13.
Peritoneal dialysis outflow failure caused by omental wrapping is a serious complication that is difficult to diagnose noninvasively. We report a case of outflow failure in which catheterography showed several characteristics of omental wrapping: "pseudocele" formation, delayed catheter emptying, and uneven distribution of contrast in the abdominal cavity. A laparoscopic procedure was performed with identification of peritoneal adhesion and greater omental wrapping. The catheter was stripped from the omentum and repositioned in the Douglas peritoneal sac. Catheterography is a reliable, safe, noninvasive, and inexpensive alternative procedure to diagnose outflow failure caused by omental wrapping.
网膜包裹导致的腹膜透析外流出路失败是一种严重的并发症,难以进行无创诊断。我们报告了一例外流出路失败的病例,其导管造影显示了网膜包裹的几个特征:“假性囊肿”形成、导管排空延迟以及腹腔内对比剂分布不均匀。通过腹腔镜手术发现了腹膜粘连和大网膜包裹。将导管从网膜上剥离并重新置于道格拉斯腹膜囊内。导管造影是一种可靠、安全、无创且经济的替代方法,可用于诊断由网膜包裹引起的外流出路失败。