Institute of Environmental Engineering of Polish Academy of Sciences, Zabrze, Poland.
J Environ Sci Health A Tox Hazard Subst Environ Eng. 2010 Aug;45(10):1275-80. doi: 10.1080/10934529.2010.493821.
This paper describes the application of the method of post-column derivatization with triiodide and UV detection at 352 nm for the determination of bromate in drinking water, mineral water and swimming pool water samples. Optimized analytical conditions were further validated in terms of accuracy, precision, linearity, limit of detection and limit of quantification. The method detection limit was at the level of 0.4 μg/L, and the spiked recovery for bromate was in the range of 92% - 104%. The method did not need a special sample treatment and was successfully applied to the analysis of bromate at the required value that is below 2.5 μg/L.
本文描述了柱后衍生化结合三碘化物和 352nm 紫外检测法在饮用水、矿泉水和游泳池水样中测定溴酸盐的应用。进一步对分析条件的准确性、精密度、线性、检测限和定量限进行了优化验证。该方法的检测限达到了 0.4μg/L,溴酸盐的加标回收率在 92%-104%范围内。该方法不需要特殊的样品处理,成功地应用于分析溴酸盐,其含量低于 2.5μg/L。