Sousa Pedro Jerónimo, Cavaco Diogo, Neves José Pedro, Adragão Pedro, Brito João, Carmo Pedro, Bouzin Martha, Bernardo Ricardo, Santos Katia Reis, Morgado Francisco Bello, Silva José Aniceto
Servigo de Cardiologia, Hospital de Santa Cruz, Centro Hospitalar de Lisboa Ocidental, Carnaxide, Portugal.
Rev Port Cardiol. 2010 Mar;29(3):451-7.
An 18-year-old patient, with hypertrophic cardiomyopathy and an implantable cardioverter defibrillator for 6 years, was admitted with inappropriate shocks as a consequence of a lead fracture (Sprint Quattro, Medtronic). The device was explanted and replaced with a new subcutaneous defibrillator (Cameron SQ-RX), without complications. This was the first time that a subcutaneous cardioverter defibrillator had been used in the Iberian Peninsula. The new implantable defibrillator, with subcutaneous lead and generator, can lower the risk of complications, including lead fracture or infection. Furthermore, this device has good rhythm diagnostic performance and therapeutic efficacy. Following the case report, we present a brief review of the new defibrillator with subcutaneous implantation.
一名18岁的肥厚型心肌病患者,植入植入式心律转复除颤器已6年,因导线断裂(美敦力Sprint Quattro)出现不适当电击而入院。该装置被取出,并用新的皮下除颤器(Cameron SQ-RX)替换,未出现并发症。这是伊比利亚半岛首次使用皮下除颤器。这种新型植入式除颤器带有皮下导线和发生器,可降低包括导线断裂或感染在内的并发症风险。此外,该装置具有良好的心律诊断性能和治疗效果。在病例报告之后,我们对新型皮下植入式除颤器进行简要综述。