J Prosthet Dent. 1991 Apr;65(4):483-7. doi: 10.1016/0022-3913(91)90284-4.
This study compared the performance of a new computerized occlusal analysis (T-Scan) system with that of Accufilm and Shimstock foil for the registration of tooth contacts on a laboratory model. Induced interceptive contacts were created on epoxy models in a mechanical closing device and the occlusal contacts were recorded for each registration method. The traditional methods were similar to the computerized analysis but the T-Scan system provided additional differential diagnostic information in force and time modes for an improved occlusal analysis.
本研究比较了一种新型计算机化咬合分析(T-Scan)系统与Accufilm和Shimstock箔片在实验室模型上记录牙齿接触情况的性能。在机械闭合装置中的环氧树脂模型上创建诱导性截获接触,并记录每种记录方法的咬合接触情况。传统方法与计算机化分析相似,但T-Scan系统在力和时间模式下提供了额外的鉴别诊断信息,以改进咬合分析。