Suppr超能文献

一种用于无框架立体定向放置脑室导管的新工具。

A new tool for frameless stereotactic placement of ventricular catheters.

机构信息

Department of Neurosurgery, International Neuroscience Institute Hannover, Hannover, Germany.

出版信息

Neurosurgery. 2010 Sep;67(3 Suppl Operative):ons131-5; discussion ons135. doi: 10.1227/01.NEU.0000382964.72262.9A.

Abstract

BACKGROUND

The accurate position of the ventricular catheter inside the frontal horn of the lateral ventricle is essential to prevent proximal failure in shunt surgery. For optimal placement, endoscopic- and image-guided techniques are available.

OBJECTIVE

We introduce a newly constructed tool for quick and safe placement of ventricular catheters. It is mounted on a fixation device and therefore allows the surgeon's optimal concentration on the catheter insertion and feeling for the penetrated tissue. To check the feasibility of the new device, we performed a study with 4 patients.

METHODS

Two patients with communicative and 2 patients with noncommunicative hydrocephalus underwent ventricular catheter placement using the new shunt placement tool. Three patients had a complex anatomy of the ventricular system.

RESULTS

In all 4 procedures, correct placement of the ventricular catheters was achieved. The additional time needed for preparations did not exceed 15 minutes. The comparison of the postoperative computed tomography scans with the preoperative planning showed good accuracy of the instrument with a mean deviation of the catheter tips from the planned position of 1.5 mm (range 1.0-2.1 mm).

CONCLUSION

The new tool allows safe and quick placement of ventricular catheters. The adjustment of the tool to the planned trajectory is performed before catheter insertion and allows optimal concentration on the insertion procedure and the fingertip feeling for the penetrated tissue.

摘要

背景

为了防止分流手术中近端故障,准确放置侧脑室额角内的脑室导管至关重要。为此,有内镜和影像引导技术可供选择。

目的

我们介绍了一种新构建的工具,用于快速安全地放置脑室导管。它安装在固定装置上,因此可以让外科医生专注于导管插入,并感受穿透的组织。为了检查新设备的可行性,我们对 4 名患者进行了一项研究。

方法

2 名交通性脑积水患者和 2 名非交通性脑积水患者使用新的分流放置工具进行脑室导管放置。其中 3 名患者的脑室系统解剖结构复杂。

结果

在所有 4 例手术中,均成功放置了脑室导管。准备工作所需的额外时间不超过 15 分钟。术后 CT 扫描与术前规划的比较显示,该器械具有良好的准确性,导管尖端与计划位置的平均偏差为 1.5 毫米(范围 1.0-2.1 毫米)。

结论

新工具允许安全、快速地放置脑室导管。在插入导管之前,可对工具进行调整,以适应计划的轨迹,从而可以专注于插入过程,并通过指尖感受穿透的组织。