Department of Pathology and Immunology, Washington University School of Medicine, St Louis, MO 63110, USA.
Am J Clin Pathol. 2010 Sep;134(3):420-8. doi: 10.1309/AJCPWEUIM2CWUOV8.
The lamellar body count (LBC) represents an alternative method to the TDx-FLM II (Abbott Laboratories, Abbott Park, IL), which is planned to be discontinued, for assessing fetal lung maturity. Our objective was to validate the LBC on 3 hematology analyzers (Coulter LH 750 and Coulter Ac.T diff2, Beckman Coulter, Brea, CA; and Sysmex XE-2100, Sysmex, Mundelein, IL) to serve as a template for other laboratories attempting to perform in-house validation. Intra-assay and interassay coefficients of variation ranged from 1.7% to 21.8% and 1.9% to 7.1%, respectively, and all analyzers demonstrated excellent linearity. Whole blood and meconium were shown to interfere with LBCs, and specimens with these contaminants should be tested using phosphatidyl glycerol. With a TDx-FLM II cutoff of 55 mg/g or more and an LBC cutoff of 50,000/microL or more for maturity, concordance between the TDx-FLM II and the LBC on all instruments was poor (<80% in all cases). Concordance between hematology analyzers was excellent (>or=94%). When laboratories are performing in-house validations, they should not correlate LBC with TDx-FLM II results without outcome data. Correlation with another validated LBC method is preferred.
板层小体计数 (LBC) 是评估胎儿肺成熟度的替代方法,替代了计划停产的 TDx-FLM II(雅培实验室,雅培公园,IL)。我们的目的是在 3 台血液学分析仪(库尔特 LH 750 和库尔特 Ac.T diff2,贝克曼库尔特,布雷亚,CA;和希森美康 XE-2100,希森美康,芒廷莱因,IL)上验证 LBC,为其他试图进行内部验证的实验室提供模板。日内和日间变异系数分别为 1.7%至 21.8%和 1.9%至 7.1%,所有分析仪均表现出出色的线性。全血和胎粪被证明会干扰 LBC,应使用磷脂甘油检测含有这些污染物的标本。当 TDx-FLM II 截断值为 55mg/g 或更高,LBC 截断值为 50,000/microL 或更高时,所有仪器上 TDx-FLM II 和 LBC 之间的一致性较差(所有情况下均<80%)。血液学分析仪之间的一致性非常好(>or=94%)。当实验室进行内部验证时,他们不应该在没有结果数据的情况下将 LBC 与 TDx-FLM II 结果相关联。与另一种经过验证的 LBC 方法的相关性更好。