Suppr超能文献

简化毛细管电泳纳米喷雾鞘流接口,用于提高肽分析的效率和灵敏度。

Simplified capillary electrophoresis nanospray sheath-flow interface for high efficiency and sensitive peptide analysis.

机构信息

Department of Chemistry, University of Washington, Seattle, WA 98195-1700, USA.

出版信息

Rapid Commun Mass Spectrom. 2010 Sep 15;24(17):2554-60. doi: 10.1002/rcm.4672.

Abstract

We report a simple nanospray sheath-flow interface for capillary electrophoresis. This interface relies on electrokinetic flow to drive both the separation and the electrospray; no mechanical pump is used for the sheath flow. This system was interfaced with an LCQ mass spectrometer. The best results were observed with a 2-microm diameter emitter tip and a 1-mm spacing between the separation capillary tip and the emitter tip. Under these conditions, mass detection limits (3sigma) of 100 amol were obtained for insulin receptor fragment 1142-1153. The separation efficiency exceeded 200,000 plates for this compound. The relative standard deviation generated during continual infusion of a 50 microM solution of angiotensin II was 2% for the total ion count and 3% for the extracted ion count over a 40-min period. Finally, the interface was also demonstrated for negative ion mode.

摘要

我们报告了一种用于毛细管电泳的简单纳米喷雾鞘流接口。该接口依赖电动流来驱动分离和电喷雾;不需要机械泵来提供鞘流。该系统与 LCQ 质谱仪相连接。在使用 2 微米直径的喷嘴和分离毛细管尖端与喷嘴之间 1 毫米的间隔的情况下,观察到最佳结果。在这些条件下,获得了胰岛素受体片段 1142-1153 的 100 amol 的质量检测限(3sigma)。对于该化合物,分离效率超过 200,000 板。在 40 分钟的时间内,连续注入 50 microM 血管紧张素 II 溶液期间产生的相对标准偏差为总离子计数的 2%,提取离子计数的 3%。最后,还证明了该接口也适用于负离子模式。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验