Suppr超能文献

[基于定量比较法的磁共振成像射频线圈图形性能研究新方法]

[A new approach to research graphical performance of MRI RF coil based on the quantitative comparison method].

作者信息

Zhu Gaojie, Li Bin, Wei Xiaoer

机构信息

The sixth people's hospital affiliated to Shanghai Jiaotong University, BME Department, Shanghai, 200233.

出版信息

Zhongguo Yi Liao Qi Xie Za Zhi. 2010 May;34(3):180-2.

Abstract

OBJECTIVE

To compare the performance of three different types of RF coil in 3.0T MRI, the signal-to-noise ratio (SNR) and image uniformity in 3.0 Tesla magnetic resonance imaging system have been studied.

MATERIALS AND METHODS

In addition to the measurement of SNR and Uniformity based on the protocol recommend by AAPM, the images had also been analyzed using SPSS and MATLAB to get the SNR distribution along two vertical direction and the whole imaging plane.

RESULTS

The SNR with T/R birdcage coil and 8-ch surface coil is seven and twelve times higher than surface coil in the region of interest (ROI), which was set in the center of the phantom images. With regard to the spatial variation of sensitivity, the uniformity of T/R birdcage coil and 8-ch surface coil are also much higher than that of surface coil.

CONCLUSION

The images from T/R birdcage and phased array coils are all better than those of surface coil. The SNR of the images from phased array coil is highest while the uniformity of the images from T/R birdcage coil is highest.

摘要

目的

为比较三种不同类型的射频线圈在3.0T磁共振成像(MRI)中的性能,研究了3.0特斯拉磁共振成像系统中的信噪比(SNR)和图像均匀性。

材料与方法

除了根据美国医学物理师协会(AAPM)推荐的协议测量SNR和均匀性外,还使用SPSS和MATLAB对图像进行分析,以获取沿两个垂直方向和整个成像平面的SNR分布。

结果

在体模图像中心设置的感兴趣区域(ROI)中,T/R鸟笼线圈和8通道表面线圈的SNR分别比表面线圈高7倍和12倍。关于灵敏度的空间变化,T/R鸟笼线圈和8通道表面线圈的均匀性也远高于表面线圈。

结论

T/R鸟笼线圈和相控阵线圈的图像均优于表面线圈。相控阵线圈图像的SNR最高,而T/R鸟笼线圈图像的均匀性最高。