反转时间和饱和时间对T1加权磁共振图像中造影剂浓度与信号强度之间关系的影响。

Effects of inversion and saturation times on relationships between contrast agent concentrations and signal intensities of T1-weighted magnetic resonance images.

作者信息

Nazarpoor Mahmood

机构信息

Department of Radiology, Faculty of Paramedicine, Tabriz University of Medical Sciences, Tabriz, Iran.

出版信息

Radiol Phys Technol. 2010 Jul;3(2):120-6. doi: 10.1007/s12194-010-0087-9. Epub 2010 Feb 27.

Abstract

The present study was an attempt to investigate the effect of variation of inversion time (T (I)) and saturation time (T (S)) on the linear relationship between contrast agent concentration and signal intensity (SI) on Turbo Fast Low Angle Shot (TurboFLASH) T (1)-weighted images in MRI. For this purpose, inversion recovery (IR) and saturation recovery (SR) sequences (Center out Phase-Encoding acquisition) were used. A phantom was designed to hold 25 vials which contained either different (between 0 and 19.77 mmol/L) or constant (1.20 mmol/L) concentrations of contrast agent. The vials of constant concentration were used for the measurement of coil non-uniformity, which was normalized to give a correction factor. The vials of different concentrations were used to measure the SI by using different sequences and different T (I) and T (S) values. To calculate the corrected SI for different concentrations, we multiplied the SI of each vial by its correction factor. The relationships between the corrected SI and the concentration [were evaluated], where the threshold of (R (2) = 0.95 and 0.99) was maintained. This study shows that different sequences and different T (I) and T (S) values can have an effect on the correlation between the SI and concentration. Regardless of the values of T (I), T (S), and the different IR and SR sequences chosen, the linear relationship between the SI and concentration was about twice that previously reported (i.e., 0.8 mmol/L, R (2) = 0.95).

摘要

本研究旨在探讨反转时间(T(I))和饱和时间(T(S))的变化对磁共振成像(MRI)中Turbo快速低角度激发(TurboFLASH)T(1)加权图像上造影剂浓度与信号强度(SI)之间线性关系的影响。为此,使用了反转恢复(IR)和饱和恢复(SR)序列(中心向外相位编码采集)。设计了一个模型来容纳25个小瓶,其中包含不同(0至19.77 mmol/L之间)或恒定(1.20 mmol/L)浓度的造影剂。恒定浓度的小瓶用于测量线圈不均匀性,并进行归一化以得到校正因子。不同浓度的小瓶用于通过使用不同序列以及不同的T(I)和T(S)值来测量SI。为了计算不同浓度下的校正SI,我们将每个小瓶的SI乘以其校正因子。评估校正后的SI与浓度之间的关系,保持(R(2)= 0.95和0.99)的阈值。本研究表明,不同序列以及不同的T(I)和T(S)值会对SI与浓度之间的相关性产生影响。无论T(I)、T(S)的值以及所选择的不同IR和SR序列如何,SI与浓度之间的线性关系约为先前报道的两倍(即0.8 mmol/L,R(2)= 0.95)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验